Diagnostics and troubleshooting
Solitrend MMP60
18
Hauser
It is also possible to perform your own calibrations and overwrite an existing calibration curve.
SD02333M
Remote display
- Description of operation and material calibration.
8.5
Special functions
Special functions that are available are described in the Operating Instructions pertaining to
the device.
9
Diagnostics and troubleshooting
The device is normally precalibrated with calibration set B and Cal.14 (air/water 0 to 100 %)
upon delivery.
Fine-tuning to achieve an accuracy of ±0.1 % in relation to the laboratory value can be
performed via a PLC or the remote display (optional).
Fine-tuning with the PLC
Depending on the PLC, it is possible to perform a parallel shift/offset in the PLC. The
parameter has different names depending on the PLC (e.g. initial load, zero point, offset,
measuring range, etc.).
‣
Perform a parallel shift / offset in the PLC
Contact the PLC manufacturer
Fine-tuning with the remote display
‣
Perform fine-tuning / parallel shift in the device via the
Offset
parameter
9.1
Different moisture value
If the moisture value of the device deviates by more than ±1 % from the laboratory value
during initial commissioning, this may be due to the following:
Incorrect installation in the material flow
The measurement surface must be sufficiently covered. A good, stable flow of material
must
be guaranteed.
‣
Correct the installation or material flow
A video of the material flow during the batch process can be helpful for analysis
purposes.
The wrong calibration curve is set
The device is delivered with calibration curve Cal.14 (air / water 0 to 100 %).
‣
Select a suitable calibration curve.