Electrical connection
Smart System for Aquacultures SSP200B
20
Hauser
1
2
3
4
5
A0025812
16
Protective ground or grounding connection
NOTICE
Protective ground or ground cable with wire ferrule or open cable lug
The cable can become loose. Loss of the protective function!
‣
To connect the protective ground or ground cable to the threaded bolt, only use a cable
with a closed cable lug as per DIN 46211, 46225, form A.
‣
Never connect the protective ground or ground cable to the threaded bolt with a wire
ferrule or an open cable lug.
6.4.3
Connecting the sensors
All sensors are supplied either with a sensor cable with an M12 connector or with a
measuring cable CYA10 with an M12 connector. Connect the cable with the M12
connector to the M12 sensor socket provided on the underside of the device. The
transmitter is already wired at the factory.
85
86
85
1
2
86
97
88
87
98
Sensor 1
PK
GY
GN
YE
BN
WH
1
A0018019
17
M12 connection
1
Sensor cable with M12 connector or measuring cable CYA10 with M12 connector, depending on the sensor
Name of connection ports: → 18
SSP200B for freshwater applications: connecting the sensors
1. Connect the sensor cable of oxygen sensor COS61D to
connection port 1
.
Содержание Smart System SSP200B
Страница 30: ...Nameplate Smart System for Aquacultures SSP200B 30 Endress Hauser 8 Nameplate ...
Страница 35: ......
Страница 36: ...www addresses endress com ...