Prosonic M FMU44 HART
Mounting
Hauser
9
Place of manufacture: See nameplate.
4.4
Storage and transport
• Pack the measuring device so that it is protected against impact.
The original packaging offers the best protection
• Permitted storage temperature: –40 to +80 °C (–40 to 176 °F)
4.4.1
Transporting the product to the measuring point
L
CAUTION
The housing or flange can be damaged.
Risk of injury!
‣
Transport the measuring device to the measuring point in its original packaging or by
the process connection.
‣
Always secure lifting equipment (belts, lugs, etc.) at the process connection, while
paying attention to the center of gravity of the device to ensure that the device does not
tilt or slip.
‣
Follow the safety instructions and transport conditions for devices weighing more than
18 kg (39.6 lb)(IEC 61010).
5
Mounting
5.1
Mounting requirements
5.1.1
Installation conditions for the sensor for level measurements
3
4
1
5
6
D
r
7
1
6
/ !
2
!
A0038210
4
Installation requirements
1
Distance to the vessel wall: ¹⁄₆ of the vessel diameter
2
Use of a weather protection cover; protection from direct sunlight or rain
3
Do not install the sensor in the middle of the vessel
4
Avoid measurements through the filling curtain
5
Do not install point level switches or temperature sensors within the beam angle
6
Symmetrical internal fixtures, e.g. heating coils and baffles, negatively impact the measurement
7
Align the sensor so that it is perpendicular to the product surface