Prosonic M FMU44 HART
Electrical connection
Hauser
17
6.2.2
Connection in the T12 housing
1 →
2 →
3 →
1.
2.
A0038256
A0038257
A0038258
‣
Unscrew the cover
‣
Avoid moisture in the housing,
provide a loop to allow
moisture to drain off
4 →
5
A0038711
A0047500
‣
Connect the cable shield to the
ground terminal in the
connection compartment
‣
Tighten the cable gland.
6. Close the housing.
7. Switch on the power supply.
6.3
Terminal assignment
2
4
3
1
2
3
4
L-
L+
I+
I-
1
A0019269
13
Terminal assignment for two-wire version
1
Supply voltage
2
Test terminal to test the signal current
3
PAL (potential equalization)
4
4-20 mA HART
‣
Connect the connecting line to the screw terminals in the connection compartment
(conductor cross-section 0.5 to 2.5 mm
2
, 20 to 14 AWG)
Содержание Prosonic M FMU44
Страница 52: ...www addresses endress com 71574097 71574097 ...