Operation
Proline Prowirl F 200 PROFIBUS PA
132
Hauser
Parameter overview with brief description
Parameter
Prerequisite
Description
User interface
Volume flow
–
Displays the volume flow currently
measured.
Dependency
The unit is taken from the
Volume flow
unit
Signed floating-point
number
Corrected volume flow
–
Displays the corrected volume flow
currently calculated.
Dependency
The unit is taken from the
Corrected
volume flow unit
parameter
Signed floating-point
number
Mass flow
–
Displays the mass flow currently
calculated.
Dependency
The unit is taken from the
Mass flow
unit
Signed floating-point
number
Flow velocity
–
Displays the flow velocity currently
calculated.
Dependency
The unit is taken from the
Velocity
unit
Signed floating-point
number
Temperature
–
Displays the temperature currently
measured.
Dependency
The unit is taken from the
Temperature unit
parameter
Signed floating-point
number
Calculated saturated steam pressure The following conditions are met:
• Order code for "Sensor version",
option "Mass (integrated
temperature measurement)"
• The
Steam
option is selected in the
Select medium
parameter
Displays the saturated steam pressure
currently calculated.
Dependency
The unit is taken from the
Pressure
unit
Signed floating-point
number
Steam quality
The following conditions are met:
• Order code for "Sensor version",
option "Mass (integrated
temperature measurement)"
• The
Steam
option is selected in the
Select medium
parameter.
Displays the current steam quality.
Dependency
Depends on the compensation mode of
the steam quality:
Steam quality
Signed floating-point
number
Total mass flow
The following conditions are met:
• Order code for "Application package",
option
EU
"Wet steam measurement"
• The
Steam
option is selected in the
Select medium
parameter
Displays the total mass flow currently
calculated (steam and condensate).
Dependency
The unit is taken from the
Mass flow
unit
Signed floating-point
number
Condensate mass flow
The following conditions are met:
• Order code for "Application package",
option
EU
"Wet steam measurement"
• The
Steam
option is selected in the
Select medium
parameter
Displays the condensate mass flow
currently calculated.
Dependency
The unit is taken from the
Mass flow
unit
Signed floating-point
number
Energy flow
With order code for "Sensor version":
option "Mass (integrated temperature
measurement)"
Displays the energy flow currently
calculated.
Dependency
The unit is taken from the
Energy flow
unit
Signed floating-point
number