Proline Promass S 100
8
Hauser
Measuring range
Measuring ranges for liquids
DN
Measuring range full scale values
min(F)
to
max(F)
[mm]
[in]
[kg/h]
[lb/min]
8
³⁄₈
0 to 2
000
0 to 73.50
15
½
0 to 6
500
0 to 238.9
25
1
0 to 18
000
0 to 661.5
40
1½
0 to 45
000
0 to 1
654
50
2
0 to 70
000
0 to 2
573
Measuring ranges for gases
The full scale values depend on the density of the gas and can be calculated with the formula below:
max(G)
=
max(F)
· ρ
G
: x
max(G)
Maximum full scale value for gas [kg/h]
max(F)
Maximum full scale value for liquid [kg/h]
max(G)
<
max(F)
max(G)
can never be greater than
max(F)
ρ
G
Gas density in [kg/m³] at operating conditions
DN
x
[mm]
[in]
[kg/m
3
]
8
³⁄₈
60
15
½
80
25
1
90
40
1½
90
50
2
90
To calculate the measuring range, use the
Applicator
sizing tool (→ 74)
Calculation example for gas
• Sensor: Promass S, DN 50
• Gas: Air with a density of 60.3 kg/m³ (at 20 °C and 50 bar)
• Measuring range (liquid):70
000 kg/h
• x = 90 kg/m³ (for Promass S, DN 50)
Maximum possible full scale value:
max(G)
=
max(F)
· ρ
G
: x = 70
000 kg/h · 60.3 kg/m³ : 90 kg/m³ = 46
900 kg/h
Recommended measuring range
Operable flow range
Over 1000 : 1.
Flow rates above the preset full scale value are not overridden by the electronics unit, with the result
that the totalizer values are registered correctly.