Содержание Proline Promass S 100

Страница 1: ...smitter housing High ingress protection IP69K Local display available Your benefits Low installation costs compact horizontal installation thanks to fully self draining tube Fewer process measuring po...

Страница 2: ...nting Safety Barrier Promass 100 40 Environment 40 Ambient temperature range 40 Storage temperature 42 Climate class 42 Degree of protection 42 Shock resistance 42 Vibration resistance 42 Interior cle...

Страница 3: ...Proline Promass S 100 Endress Hauser 3 Registered trademarks 76...

Страница 4: ...hing any other connections Equipotential connection A connection that has to be connected to the plant grounding system This may be a potential equalization line or a star grounding system depending o...

Страница 5: ...ow causes deceleration of the oscillation at the inlet of the tubes 2 and acceleration at the outlet 3 1 2 3 A0016772 The phase difference A B increases with increasing mass flow Electrodynamic sensor...

Страница 6: ...rosoft Internet Explorer Also for device version with EtherNet IP output Via Web browser e g Microsoft Internet Explorer Via Add on Profile Level 3 for automation system from Rockwell Automation Via E...

Страница 7: ...nd Zone 1 Div 1 Safety IT security We only provide a warranty if the device is installed and used as described in the Operating Instructions The device is equipped with security mechanisms to protect...

Страница 8: ...max G max F max G can never be greater than max F G Gas density in kg m at operating conditions DN x mm in kg m3 8 60 15 80 25 1 90 40 1 90 50 2 90 To calculate the measuring range use the Applicator...

Страница 9: ...tput Version Passive open collector Maximum input values DC 30 V 25 mA Voltage drop For 25 mA DC 2 V Pulse output Pulse width Adjustable 0 05 to 2000 ms Maximum pulse rate 10000 Impulse s Pulse value...

Страница 10: ...EIA TIA 485 A standard Terminating resistor For device version used in non hazardous areas or Zone 2 Div 2 integrated and can be activated via DIP switches on the transmitter electronics module For d...

Страница 11: ...odbus RS485 Failure mode Choose from NaN value instead of current value Last valid value EtherNet IP Device diagnostics Device condition can be read out in Input Assembly Local display Plain text disp...

Страница 12: ...A 72 B Uo 16 24 V Io 623 mA Po 2 45 W For IIC Lo 92 8 H Co 0 433 F Lo Ro 14 6 H For IIB Lo 372 H Co 2 57 F Lo Ro 58 3 H The gas group depends on the sensor and nominal diameter For an overview and for...

Страница 13: ...c variable Mass flow Volume flow Corrected volume flow Density Reference density Temperature Totalizer 1 Totalizer 2 Totalizer 3 The range of options increases if the measuring device has one or more...

Страница 14: ...ectronic temperature Oscillation frequency Oscillation amplitude Frequency fluctuation Oscillation damping Tube damping fluctuation Signal asymmetry Exciter current Digital input 1 to 2 Partially fill...

Страница 15: ...s 16 Write multiple registers 23 Read write multiple registers Broadcast messages Supported by the following function codes 06 Write single registers 16 Write multiple registers 23 Read write multiple...

Страница 16: ...evel Ring DLR No Fix Input RPI 5 ms to 10 s factory setting 20 ms Exclusive Owner Multicast Instance Size byte Instance configuration 0x68 398 O T configuration 0x66 64 T O configuration 0x64 44 Exclu...

Страница 17: ...asuring device has one or more application packages Fix Output Output Assembly Activation of reset totalizers 1 3 Activation of pressure compensation Activation of reference density compensation Activ...

Страница 18: ...ug 1 Connection version 4 20 mA HART pulse frequency switch output 1 1 Signal transmission pulse frequency switch output 1 2 Signal transmission 4 20 mA HART 1 3 Supply voltage 2 Connection version Mo...

Страница 19: ...G Option U plug M12x1 thread M20 Options A B C Device plug Device plug Option Q 2 x plug M12x1 Order code for Housing Option A compact coated aluminum Option B compact hygienic stainless Option C ult...

Страница 20: ...ad NPT Options A B Device plug Terminals Option L plug M12x1 thread NPT Option N plug M12x1 coupling M20 Option P plug M12x1 thread G Option U plug M12x1 thread M20 Options A B C Device plug Device pl...

Страница 21: ...12x1 thread NPT Option N plug M12x1 coupling M20 Option P plug M12x1 thread G Option U plug M12x1 thread M20 Options A B C Device plug Device plug Option Q 2 x plug M12x1 Order code for Housing Option...

Страница 22: ...1 Option C thread G Option D thread NPT A B C Device plug Option I plug M12x1 Order code for Housing Option A compact coated aluminum Option B compact hygienic stainless Option C ultra compact hygieni...

Страница 23: ...M12x1 thread NPT Option N plug M12x1 coupling M20 Option P plug M12x1 thread G Option U plug M12x1 thread M20 Options A B C Device plug Device plug Option Q 2 x plug M12x1 Order code for Housing Optio...

Страница 24: ...4 20 mA HART pulse frequency switch output 18 PROFIBUS DP 20 Modbus RS485 21 EtherNet IP 23 Supply voltage For all connection versions except MODBUS RS485 intrinsically safe device side Device plug MO...

Страница 25: ...ce plug for signal transmission device side 3 2 4 1 5 A0016811 Pin Assignment Coding Plug socket 1 B Socket 2 A PROFIBUS DP 3 4 B PROFIBUS DP 5 Grounding shielding Recommended plug Binder series 763 p...

Страница 26: ...99 3729 810 04 Phoenix part no 1543223 SACC M12MSD 4Q When using the device in a hazardous location Use a suitably certified plug Supply voltage Transmitter For device version with all communication...

Страница 27: ...M Modbus RS485 for use in intrinsically safe areas 230 mA 10 A 0 8 ms Power supply failure Totalizers stop at the last value measured Depending on the device version the configuration is retained in t...

Страница 28: ...67 4 Resistor for HART communication 250 observe maximum load 9 5 Analog display unit observe maximum load 9 6 Transmitter Pulse frequency output 1 _ 12345 2 3 A0016801 9 Connection example for pulse...

Страница 29: ...specifications 32 3 Distribution box 4 Transmitter If baud rates 1 5 MBaud an EMC cable entry must be used and the cable shield must continue as far as the terminal wherever possible Modbus RS485 2 1...

Страница 30: ...able shield observe cable specifications 32 3 Safety Barrier Promass 100 4 Observe cable specifications 32 5 Non hazardous area 6 Non hazardous area and Zone 2 Div 2 7 Intrinsically safe area 8 Transm...

Страница 31: ...current input e g PLC Prerequisite automation system with HART version 6 HART commands 113 and 114 can be processed 2 Cable shield observe cable specifications 32 3 Resistor for HART communication 25...

Страница 32: ...34 mm2 22 AWG Cable type Twisted pairs Loop resistance 110 km Signal damping Max 9 dB over the entire length of the cable cross section Shielding Copper braided shielding or braided shielding with fo...

Страница 33: ...m cable length mm2 AWG m ft 0 5 20 70 230 0 75 18 100 328 1 0 17 100 328 1 5 16 200 656 2 5 14 300 984 Performance characteristics Reference operating conditions Error limits based on ISO 11631 Water...

Страница 34: ...0 700 140 US units DN 1 1 1 10 1 20 1 50 1 100 1 500 inch lb min lb min lb min lb min lb min lb min 73 50 7 350 3 675 1 470 0 735 0 147 238 9 23 89 11 95 4 778 2 389 0 478 1 661 5 66 15 33 08 13 23 6...

Страница 35: ...e is a difference between the temperature for zero point adjustment and the process temperature the typical measured error of the sensor is 0 0002 of the full scale value C 0 0001 of the full scale va...

Страница 36: ...Value measured value ZeroPoint zero point stability Calculation of the maximum measured error as a function of the flow rate Flow rate Maximum measured error in o r ZeroPoint BaseAccu 100 A0021332 Bas...

Страница 37: ...ipe Highest point of a pipeline Directly upstream of a free pipe outlet in a down pipe A0023344 Installation in down pipes However the following installation suggestion allows for installation in an o...

Страница 38: ...may reduce the ambient temperature To maintain the minimum ambient temperature for the transmitter this orientation is recommended 2 Applications with high process temperatures may increase the ambien...

Страница 39: ...mm 298 402 542 750 1019 B mm 33 33 33 36 5 44 1 C mm 28 28 38 56 75 US units DN in 1 1 2 A in 11 73 15 83 21 34 29 53 40 12 B in 1 3 1 3 1 3 1 44 1 74 C in 1 1 1 1 1 5 2 2 2 95 Zero point adjustment A...

Страница 40: ...fety Barrier Promass 100 40 to 60 C 40 to 140 F If operating outdoors Avoid direct sunlight particularly in warm climatic regions Weather protection covers can be ordered from Endress Hauser see Acces...

Страница 41: ...a maximum medium temperature Tm 392 F Tm 392 F Ex nA CCSAUS NI SI units Order code for Housing Ta C T6 85 C T5 100 C T4 135 C T3 200 C T2 300 C T1 450 C Option A Compact coated alu Option B Compact hy...

Страница 42: ...is immediately greater than or equal to the measured maximum ambient temperature Tma that is present Ta 50 C The row showing the maximum medium temperature is determined 3 Select the maximum medium te...

Страница 43: ...cable shield must continue as far as the terminal wherever possible For details refer to the Declaration of Conformity Process Medium temperature range Sensor 50 to 150 C 58 to 302 F Seals No internal...

Страница 44: ...50 0 50 C 200 150 100 50 0 50 400 0 10 20 30 40 bar psi 200 400 600 0 A0020971 EN 23 With flange material 1 4404 316 316L Process connection to DIN 11851 300 F 350 250 200 150 100 50 0 50 C 200 150 10...

Страница 45: ...DIN 11864 2 Form A DN 8 40 DN 50 300 F 350 250 200 150 100 50 0 50 C 200 150 100 50 0 50 400 0 10 20 30 40 bar psi 200 400 600 0 A0020968 EN 26 With flange material 1 4404 316L Process connection to...

Страница 46: ...cess characteristics e g with corrosive process fluids we recommend the use of sensors whose secondary containment is equipped with special pressure monitoring connections order code for Purge connect...

Страница 47: ...liquids do not outgas This is prevented by means of a sufficiently high system pressure For this reason the following mounting locations are recommended At the lowest point in a vertical pipe Downstre...

Страница 48: ...re of Ta 60 C 140 F NOTICE The insulation can also be thicker than the recommended insulation thickness Prerequisite The temperature at the lower end of the transmitter housing does not exceed 80 C 17...

Страница 49: ...191 312 17 60 121 92 3 40 136 147 5 93 5 54 215 388 26 00 173 132 3 50 136 147 5 93 5 54 226 467 40 50 241 167 3 1 If using an extension neck for the extended temperature range order code for Sensor...

Страница 50: ...Dimensions SI units DN mm A mm B mm C mm D mm E 2 mm F 1 2 mm G mm H mm K mm L mm 8 133 5 136 8 78 58 8 186 294 8 31 108 92 3 15 133 5 136 8 78 58 8 186 294 12 00 108 92 3 25 133 5 136 8 78 58 8 186...

Страница 51: ...connection Order code for Housing option C Ultra compact hygienic stainless M12 device plug A D C B E H F G K L A0019438 Dimensions SI units DN mm A mm B mm C mm D mm E 2 mm F 1 2 mm G mm H mm K mm L...

Страница 52: ...in 2 If using a display order code for Display Operation option B values 0 55 in 3 dependent on respective process connection Process connections in SI units Flange connections EN DIN C D A B E L 1 5...

Страница 53: ...s connection option AAW Surface roughness flange Ra 3 2 to 6 3 m DN mm A mm B mm C mm D mm E mm L mm 8 1 90 60 3 4 15 7 17 1 15 7 336 15 90 60 3 4 15 7 17 1 15 7 440 25 110 79 4 4 15 7 17 6 26 7 580 4...

Страница 54: ...nection option NEW Surface roughness flange Ra 3 2 to 6 3 m DN mm A mm B mm C mm D mm E mm L mm 8 1 95 70 4 15 16 0 15 336 15 95 70 4 15 16 0 15 440 25 125 90 4 19 17 5 25 580 40 140 105 4 19 20 0 40...

Страница 55: ...l option LP 1 Tri Clamp 1 4435 316L order code for Process connection option FNW DN mm Clamp in A mm B mm L mm 8 1 50 4 22 10 362 15 1 50 4 22 10 466 3A version available order code for Additional app...

Страница 56: ...L mm 8 Rd 34 16 362 15 Rd 34 16 466 25 Rd 52 26 606 40 Rd 65 38 825 50 Rd 78 50 1107 3A version available order code for Additional approval option LP Rd 28 1 8 Threaded hygienic connection DIN 11851...

Страница 57: ...IN 11864 2 Form A flange with groove C D A B E X X L 1 5 0 06 2 0 0 08 A0015627 37 Detail X Asymmetrical process connection the part shown in gray is provided by the supplier Engineering unit mm in DI...

Страница 58: ...m A 1 4435 316L order code for Process connection option KMW DN mm A mm B mm L mm 8 34 0 16 05 370 15 34 0 16 05 474 25 50 5 26 05 614 40 64 0 38 05 825 50 77 5 50 05 1096 3A version available order c...

Страница 59: ...oval option LP ISO 2852 clamp connection L 1 5 0 06 2 0 0 08 A B A0015625 40 Engineering unit mm in Clamp connection ISO 2852 1 4435 316L order code for Process connection option JSB DN mm A mm B mm L...

Страница 60: ...1107 3A version available order code for Additional approval option LP SMS 1145 threaded hygienic connection L 1 5 0 06 2 0 0 08 A B A0015628 42 Engineering unit mm in Threaded hygienic connection SM...

Страница 61: ...37 4 0 62 0 67 0 62 13 23 3 54 2 37 4 0 62 0 67 0 62 17 32 1 4 33 3 13 4 0 62 0 69 1 05 22 83 1 4 92 3 87 4 0 62 0 73 1 61 31 26 2 5 91 4 75 4 0 75 0 99 2 07 42 17 1 DN with DN flanges as standard Fla...

Страница 62: ...l option LP 1 Tri Clamp 1 4435 316L order code for Process connection option FNW DN in Clamp in A in B in L in 1 1 98 0 87 14 25 1 1 98 0 87 18 35 3A version available order code for Additional approv...

Страница 63: ...16L order code for Process connection option SAW DN in A in B in L in Rd 40 0 89 14 25 Rd 40 0 89 18 35 1 Rd 40 0 89 23 86 1 Rd 60 1 4 32 64 2 Rd 70 1 91 43 58 3A version available order code for Addi...

Страница 64: ...3 23 204 8 04 25 1 NPT 25 0 98 82 3 23 348 13 54 40 1 NPT 45 1 77 102 4 02 526 20 70 50 2 NPT 58 2 28 119 5 4 70 763 30 04 Weight Compact version Weight in SI units All values weight refer to devices...

Страница 65: ...1 5 3 Adapter for cable entry with internal thread G or NPT Order code for Housing option A Compact coated aluminum The various cable entries are suitable for hazardous and non hazardous areas Cable e...

Страница 66: ...thout internal seals Safety Barrier Promass 100 Housing Polyamide Process connections Flanges EN 1092 1 DIN 2501 EN 1092 1 DIN 2512N ASME B16 5 JIS B2220 Tri Clamp OD tubes Clamp with aseptic connecti...

Страница 67: ...Local display The local display is only available with the following device version Order code for Display Operation option B 4 line via communication Display element 4 line liquid crystal display wi...

Страница 68: ...tool with COM DTM CDI Communication TCP IP 5 Ethernet switch Via PROFIBUS DP network This communication interface is present in the following device version Order code for Output option L PROFIBUS DP...

Страница 69: ...Web server or with FieldCare operating tool with COM DTM CDI Communication TCP IP 3 Standard Ethernet connecting cable with RJ45 plug PROFIBUS DP 1 2 3 A0021270 49 Connection for order code for Output...

Страница 70: ...ion interface is present in the following device version Order code for Output option M Modbus RS485 1 2 3 A0016925 1 Service interface CDI of the measuring device 2 Commubox FXA291 3 Computer with Fi...

Страница 71: ...ng device is certified and registered by the PROFIBUS User Organization PNO The measuring system meets all the requirements of the following specifications Certified in accordance with PROFIBUS PA Pro...

Страница 72: ...EMC of industrial process and laboratory control equipment NAMUR NE 32 Data retention in the event of a power failure in field and control instruments with microprocessors NAMUR NE 43 Standardization...

Страница 73: ...ntervals in accordance with operator s risk assessment Concentration Package Description Concentration measurement and special density Calculation and outputting of fluid concentrations Many applicati...

Страница 74: ...r HART and FOUNDATION Fieldbus devices in the non Ex area For details see Operating Instructions BA01202S Field Xpert SFX370 Field Xpert SFX370 is a mobile computer for commissioning and maintenance I...

Страница 75: ...erating Instructions BA00247R iTEMP The temperature transmitters can be used in all applications and are suitable for the measurement of gases steam and liquids They can be used to read in the fluid t...

Страница 76: ...e part sets Specified for each individual accessory 73 Registered trademarks HART Registered trademark of the HART Communication Foundation Austin USA PROFIBUS Registered trademark of the PROFIBUS Use...

Страница 77: ......

Страница 78: ......

Страница 79: ......

Страница 80: ...www addresses endress com...