Proline Promass 80
Commissioning
Hauser
23
5.6 Troubleshooting
The error messages that can occur most frequently when a measuring device is commissioned
are described here.
A complete description of all the error messages
→
Operating Instructions on the CD-ROM.
General
Note! The output signals (e.g. pulse, frequency) of the measuring device must correspond to the
higher-order controller.
HART
No.
Error message / Type
Cause/remedy
351 to 354
System error message (S)/
Notice message (!)
CURRENT SPAN n
# 351 to 354
Current output
The current flow is outside the set range.
1. Change the upper range or lower range values entered
2. Increase or reduce flow, as applicable
701
Process error message (P)/
Notice message (!)
EXC. CURR. LIM
# 701
The maximum current value for the measuring tube exciter coils has
been reached since certain fluid characteristics, e.g. high gas or solid
content, are in the limit range. The device continues to work
correctly.
In particular with outgassing fluids and/or increased gas content, the
following measures are recommended to increase system pressure:
1. Install the measuring device downstream of a pump
2. Mount the device at the lowest point in an ascending pipeline
3. Install a valve or an orifice plate downstream from the
measuring device