Commissioning
Proline Promass 80
24
Hauser
PROFIBUS
No.
Device status message
(onsite display)
PROFIBUS measured value
status
Extended
diagnostic
message in
the
PROFIBUS
master
Cause/remedy
351
…
354
System error mess. (S)/
notice message (!)
CURRENT SPAN n
# 351 to 354
• Quality Code (HEX),
measured value status:
0x54; 0x55; 0x56
• Quality status:
UNCERTAIN
• Quality substatus:
Engineering unit range
violation
• Limits: O.K.; Low; High
Flow is out of
range
See HART table
701
Process error mess. (P)/
Notice message (!)
EXC. CURR. LIM
# 701
• Quality Code (HEX),
measured value status:
0x40; 0x41; 0x42
• Quality status:
UNCERTAIN
• Quality substatus:
Non-specific
• Limits: O.K.; Low; High
Excitation too
high
See HART table
Содержание Proline Promass 80