Pipes with cathodic protection
If the medium cannot be grounded for process-related reasons, the measuring
device must be installed in such a way as to be potential-free:
• When installing the measuring device, make sure that there is an electrical
connection between the two piping runs, 9 AWG copper wire (6 mm
2
).
• Make sure that the installation materials do not establish a conductive connection
to the measuring device and that the installation materials withstand the
tightening torques applied when the threaded fasteners are tightened.
• Also comply with the regulations applicable to potential-free installation.
1 = isolating transformer, 2 = electrically insulated
Performance characteristics
To DIN 19200 and VDI/VDE 2641:
• Medium temperature: +28 °C ± 2 K
• Ambient temperature: +22 °C ± 2 K
• Warm-up time: 30 minutes
Installation:
• Inlet run >10 x Dia
• Outlet run > 5 x Dia
• Sensor and transmitter grounded.
• Sensor centered relative to the pipe.
Pulse output: ± 0.5% o.r. ± 0.02% of max. full scale (o.r. = of reading)
Current output: plus typically ± 5
µ
A
Supply-voltage fluctuations have no effect within the specified range.
Reference conditions
Measured error
Repeatability
Measured error in [%] of reading
Maximum ± 0.2% o.r. ± 0.02% of maximum full scale (o.r. = of reading)