PROline Promag 10 P

Hauser

12

Conductivity

Minimum conductivity:  

 50

µ

S/cm

Note that in the case of the remote version, the minimum conductivity is also
influenced by the length of the connecting cable.  Refer to “Length of connecting
cable” on Page 11.

ANSI B 16.5:
Class 150, 1” to 12”

Medium pressure range
(nominal pressure)

As a rule, the pipeline diameter determines the sensor diameter.  With a known
flowrate and the help of the Hauser Flow Applicator, the best size can be
selected.  Generally, a velocity range of 6 to 10 ft/sec should be selected.
A necessary increase in velocity can be achieved through a reduction of the sensor
diameter.  The higher installation expense is normally balanced by the lower sensor
cost.
The flow velocity should also be determined by the fluid’s physical properties:
• v < 6 ft/sec, for abrasive mediums such as potter’s clay, lime milk, ore slurry, etc.
• v > 6 ft/sec, for fluids causing build-up such as wastewater sludge, etc.

Flow limit

Pressure tightness (liner)

Size

Measuring Resistance to partial vacuum
tube lining of measuring tube lining

Limit values for abs. pressure (psia)
at various fluid temperatures
75°F

175°F

212°F

265°F

1”

PTFE

0

0

0

1.4

1-1/2” PTFE

0

0

0

1.4

2”

PTFE

0

0

0

1.4

3”

PTFE

0

*

0.6

1.9

4”

PTFE

0

*

1.9

2.5

6”

PTFE

1.9

*

3.5

5.6

8”

PTFE

2.9

*

4.2

5.9

10”

PTFE

4.8

*

5.8

7.7

12”

PTFE

5.8

*

7.3

9.1

* No value can be specified.

Pressure loss

No pressure loss if the sensor is installed in a pipe of the same nominal diameter.
Pressure losses occur in configurations incorporating adapters (reducers, expand-
ers).  Contact your Hauser representative or Hauser for informa-
tion.

Nominal

diameter

Min./max. full scale value

(v ~ 1.0 or 33 ft/s)

Full scale value

(v ~ 8 ft/s)

Pulse weighting

(~ 2 pulses/s)

Creepage

(~ 0.1 ft/s)

inch

1

1-1/2

2
3
4
6
8

10
12

mm

25
40
50
80

100
150
200
250
300

2.5 to 80 gpm

7 to 190 gpm

10 to 300 gpm
24 to 800 gpm

40 to 1250 gpm
90 to 2650 gpm

155 to 4850 gpm
250 to 7500 gpm

350 to 10,600 gpm

18 gpm
50 gpm
75 gpm

200 gpm
300 gpm
600 gpm

1200 gpm
1500 gpm
2400 gpm

0.20 gal
0.50 gal

0.5 gal

2 gal
2 gal
5 gal

10 gal
15 gal
25 gal

0.25 gpm
0.75 gpm
1.25 gpm

2.5 gpm
4.0 gpm

12 gpm
15 gpm
30 gpm
45 gpm

Recommended

flow rate

Factory settings

Содержание PROline promag 10 P

Страница 1: ...t for driving an external totalizer or as status output Guaranteed safety of application Promag 10 offers all basic functions to guarantee a high degree of reliability and measured value stability Opt...

Страница 2: ...Q A v Ue induced voltage B magnetic induction magnetic field L electrode gap v flow velocity Q volume flow A pipe cross section I current strength The measuring system consists of a transmitter and a...

Страница 3: ...ally isolated from each other Output signal Load Signal on alarm Low flow cutoff Galvanic isolation Output Power supply Electrical connection measuring unit a Cable for power supply 85 to 250 VAC 20 t...

Страница 4: ...grounded 36 pF ft 120 pF m Permanent operating temperature 5 to 180 F 20 to 80 C Signal cable 3 x 20 AWG 0 38 mm2 PVC cable with common braided copper shield approximately 0 28 diameter approx 7 mm a...

Страница 5: ...t 28 VAC maximum 16 A 5 ms at 250 VAC Lasting minimum 1 2 power cycle EEPROM retain the measuring system data in the event of a power supply failure Power consumption Power supply failure Potential eq...

Страница 6: ...pipes and lined pipes Normally potential is matched using the reference electrodes in the measuring tube However in exceptional cases it is possible that due to the grounding plan of a system large ma...

Страница 7: ...d Also comply with the regulations applicable to potential free installation 1 isolating transformer 2 electrically insulated Performance characteristics To DIN 19200 and VDI VDE 2641 Medium temperatu...

Страница 8: ...isk of damage to the lining of the measuring tube It might be necessary to install pulse dampers in systems incorporating reciprocat ing diaphragm or peristaltic pumps The sensor is shock and vibratio...

Страница 9: ...for correct measuring of problematic mediums Horizontal orientation The measuring electrode plane should be horizontal This prevents brief insulation of the two electrodes by entrained air bubbles Cau...

Страница 10: ...tlet runs If possible install the sensor well clear of fittings such as valves T pieces elbows etc Compliance with the following requirements for the inlet and outlet runs is necessary in order to ens...

Страница 11: ...temperature 5 to 140 F 20 to 60 C for sensor and transmitter Install the device at a shady location Avoid direct sunlight particularly in warm climatic regions 15 to 120 F 10 to 50 C 70 F 20 C is the...

Страница 12: ...Measuring Resistance to partial vacuum tube lining of measuring tube lining Limit values for abs pressure psia at various fluid temperatures 75 F 175 F 212 F 265 F 1 PTFE 0 0 0 1 4 1 1 2 PTFE 0 0 0 1...

Страница 13: ...10 remote transmitter mounting A Direct wall mounting B Pipe mounting Esc E ANSCHLUSSKLEMMEN FIELD TERMINALS 4 49 113 7 01 178 5 31 135 3 94 100 3 94 100 5 91 150 4 76 121 0 78 1 18 20 30 6 34 7 13 1...

Страница 14: ...19 7 200 200 200 200 250 300 350 450 500 331 331 331 381 381 462 517 567 617 247 247 247 272 272 312 337 362 387 9 72 9 72 9 72 10 7 10 7 12 3 13 3 14 2 15 2 3 31 3 31 3 31 4 29 4 29 5 90 7 08 8 07 9...

Страница 15: ...336 336 417 472 522 572 202 202 202 227 227 267 292 317 342 7 95 7 95 7 95 8 94 8 94 10 5 11 5 12 5 13 5 A in mm B in mm C in mm K in mm E in mm Weight lb kg Nominal Size 1 1 1 2 2 3 4 6 8 10 12 25 4...

Страница 16: ...22 Seal Seals to DIN 2690 Ground disks 316L SS or Alloy C 22 0 59 15 0 39 10 Nominal size 1 0 08 2 H B A D 0 26 6 5 1 1 1 2 2 3 4 6 8 10 12 2 1 18 1 75 2 22 3 35 4 33 6 42 8 29 10 4 12 5 30 44 5 56 5...

Страница 17: ...aboratory Procedures EN 61326 IEC 1326 Emmision to class A requirements Electromagnetic compatibility EMC require ments ANSI ISA S82 01 Safety Standards for Electrical and Electronic Test Measuring Co...

Страница 18: ...Hauser are available for the transmitter and the sensor Contact your local representative or Endress Hauser for detailed information Supplemental documentation Technical Information Promag 10 W TI 09...

Страница 19: ...7 Approvals A For use in nonhazardous area R FM approved non incendive Class I Division 2 CSA Class I Division 2 8 Protection type A NEMA 4X IP 67 compact aluminum field housing G NEMA 4X IP 67 remote...

Страница 20: ...el Pedregal 01900 Mexico D F Mexico Phone 525 568 2405 FAX 525 568 7459 For application and selection assistance in the U S call 888 ENDRESS For total support of your installed base 24 hours a day in...

Отзывы: