Oxymax COS22D
Technical data
Hauser
35
10.2.13 Temperature compensation
Compensation of the membrane properties is realized in the transmitter from 0 to 90
˚
C (32 to 194
˚
F),
extrapolation from 90
˚
C (194
˚
F)
• Measured variable as partial pressure [hPa] or in Vol%: -5 to 90
˚
C (23 to 194
˚
F)
• Measured variable as concentration [mg/l]: 0 to 80
˚
C (32 to 176
˚
F)
• Measured variable as saturation [%SAT]: -5 to 90
˚
C (23 to 194
˚
F)
10.3 Environment
10.3.1 Storage temperature
–5 to +50
˚
C (20 to 120
˚
F) at 95% relative air humidity, not condensing
NOTICE
Danger of drying out
►
Only store the sensor with the protection cap (filled with tap water).
10.3.2 Ambient temperature range
–5 to +135
˚
C (23 to 175
˚
F), non-freezing
10.3.3 Ingress protection
IP 68 (10 m (33 ft) water column at 25
˚
C (77
˚
F) during 45 days, 1 mol/l KCl)
10.4 Process
10.4.1 Process temperature
–5 to +135
˚
C (23 to 175
˚
F), non-freezing
10.4.2 Process pressure
ambient pressure to 12 bar (174 psi)
a0011887
Fig. 12: Defined ranges of temperature compensation depending on the unit of measure