Maintenance
Oxymax COS22D
26
Hauser
7.3
Cleaning the sensor
The measurement can be corrupted by sensor fouling or malfunction, e.g.:
• --> cause longer response times and a reduced slope under certain circumstances.
To ensure reliable measurement, the sensor must be cleaned at regular intervals. The frequency and
intensity of the cleaning operation depend on the measuring medium.
Clean the sensor:
• before every calibration
• at regular intervals during operation as necessary
• before returning it for repairs.
Depending on the type of soiling, proceed as follows:
After cleaning, rinse the sensor with copious amounts of clean water.
7.4
Wear parts and consumables
Parts of the sensor will suffer wear and tear during operation.
Suitable action can restore normal operating functionality. This action includes:
Type of soiling
Cleaning
Salt deposits
Immerse the sensor in drinking water or in 1-5% hydrochloric acid
for a few minutes. Afterwards, rinse it with copious amounts of water.
Dirt particles on the sensor body and the
sensor cap
(not membrane!)
Clean the sensor body and cap mechanically with water and a suitable
brush.
Dirt particles on the membrane body or the
membrane
Clean with water and a soft sponge.
Action
Cause
Replacing the sealing ring
visual damage to a sealing ring
Replacing the electrolyte
unstable or implausible measuring signal or electrolyte soiling
Replacing the membrane body
uncleanable membrane, damaged membrane (hole or overstretch)
Содержание Oxymax COS22D
Страница 1: ...BA00447C 07 EN 02 12 71153853 Operating Instructions Oxymax COS22D Dissolved oxygen sensor...
Страница 7: ...Oxymax COS22D Identification Endress Hauser 7 a0014925 Fig 1 EC declaration of conformity...
Страница 16: ...Wiring Oxymax COS22D 16 Endress Hauser a0010056 Fig 8 Connection in hazardous locations...
Страница 40: ...BA00447C 07 EN 02 12 FM9...