Oxymax COS22D
Installation
Hauser
11
3.3.2
Installing a measuring point
WARNING
!
Voltage
In case of faults not-grounded, metalic assemblies may be energized
►
When using metallic assemblies and installation equipment, comply with national grounding
regulations.
For a complete installation of a measuring point, proceed as follows:
1.
Install a retractable or a flow assembly (if used) into the process.
2.
Connect the water supply to the rinse connections (if you use an assembly with cleaning function).
3.
Install and connect the oxygen sensor.
NOTICE
Installation errors
Cable breakage, loss of the sensor, unintended unscrewing of the membrane cap
►
Do not install the sensor suspended from the cable.
►
Screw the sensor into the assembly so that the cable is not twisted.
►
Mounting or dismounting the sensor: hold the sensor body, only turn the threaded plug-in head
while screwing the sensor in or out of the assembly. Otherwise you could unscrew the sensor’s
membrane cap. It will remain in the assembly resp. in the process then.
►
Avoid exerting excessive tensile force on the cable (e.g. from jerky pulling).
►
Select the installation location so that there is easy access for later calibration.
3.4
Installation examples
3.4.1
Fixed installation (CPA442)
The CPA442 fixed installation assembly allows a sensor to be easily adapted to almost any process
connections from Ingold nozzles to Varivent or Triclamp connections.
This type of installation is very suitable for tanks and larger pipes. The sensor achieves a defined
immersion depth in the medium in an easy manner.
Содержание Oxymax COS22D
Страница 1: ...BA00447C 07 EN 02 12 71153853 Operating Instructions Oxymax COS22D Dissolved oxygen sensor...
Страница 7: ...Oxymax COS22D Identification Endress Hauser 7 a0014925 Fig 1 EC declaration of conformity...
Страница 16: ...Wiring Oxymax COS22D 16 Endress Hauser a0010056 Fig 8 Connection in hazardous locations...
Страница 40: ...BA00447C 07 EN 02 12 FM9...