Oxymax COS22D
Function
Hauser
21
5.2.3
Calculation example for the calibration value
As a check, you can calculate the expected calibration value (transmitter display) as shown in the
following example (salinity is 0).
1.
Determine:
– The ambient temperature for the sensor (air temperature for "air" calibration method,water
temperature for "air-saturated water" calibration type)
– the altitude above sea level
– the current air pressure (=rel. air pressure to sea level) at the time of calibration. (If
undeterminable, use 1013 hPa (407 inH
2
O) for an approximate calculation.)
2.
Define:
– the saturation value
S
acc. to the first table
– the factor
K
acc. to the second table
T
[
˚
C (
˚
F)]
S
[mg/l=ppm]
T
[
˚
C (
˚
F)]
S
[mg/l=ppm]
T
[
˚
C (
˚
F)]
S
[mg/l=ppm]
T
[
˚
C (
˚
F)]
S
[mg/l=ppm]
0 (32)
14.64
11 (52)
10.99
21 (70)
8.90
31 (88)
7.42
1 (34)
14.23
12 (54)
10.75
22 (72)
8.73
32 (90)
7.30
2 (36)
13.83
13 (55)
10.51
23 (73)
8.57
33 (91)
7.18
3 (37)
13.45
14 (57)
10.28
24 (75)
8.41
34 (93)
7.06
4 (39)
13.09
15 (59)
10.06
25 (77)
8.25
35 (95)
6.94
5 (41)
12.75
16 (61)
9.85
26 (79)
8.11
36 (97)
6.83
6 (43)
12.42
17 (63)
9.64
27 (81)
7.96
37 (99)
6.72
7 (45)
12.11
18 (64)
9.45
28 (82)
7.82
38 (100)
6.61
8 (46)
11.81
19 (66)
9.26
29 (84)
7.69
39 (102)
6.51
9 (48)
11.53
20 (68)
9.08
30 (86)
7.55
40 (104)
6.41
10 (50)
11.25
Содержание Oxymax COS22D