LPGmass Modbus RS485
Diagnostics and troubleshooting
Hauser
63
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
083
Memory content
1. Restart device
2. Contact service
F
Alarm
140
Sensor signal
1. Check or change main
electronics
2. Change sensor
S
Alarm
144
Measuring error too high
1. Check or change sensor
2. Check process conditions
F
Alarm
190
Special event 1
Contact service
F
Alarm
191
Special event 5
Contact service
F
Alarm
192
Special event 9
Contact service
F
Alarm
Diagnostic of electronic
242
Software incompatible
1. Check software
2. Flash or change main
electronics module
F
Alarm
270
Main electronic failure
Change main electronic module F
Alarm
271
Main electronic failure
1. Restart device
2. Change main electronic
module
F
Alarm
272
Main electronic failure
1. Restart device
2. Contact service
F
Alarm
273
Main electronic failure
Change electronic
F
Alarm
274
Main electronic failure
Change electronic
S
Warning
311
Electronic failure
1. Reset device
2. Contact service
F
Alarm
390
Special event 2
Contact service
F
Alarm
391
Special event 6
Contact service
F
Alarm
392
Special event 10
Contact service
F
Alarm
Diagnostic of configuration
410
Data transfer
1. Check connection
2. Retry data transfer
F
Alarm
411
Up-/download active
Up-/download active, please
wait
C
Warning
438
Dataset
1. Check data set file
2. Check device configuration
3. Up- and download new
configuration
M
Warning
453
Flow override
Deactivate flow override
C
Warning
484
Simulation failure mode
Deactivate simulation
C
Alarm
485
Simulation measured
variable
Deactivate simulation
C
Warning
590
Special event 3
Contact service
F
Alarm
591
Special event 7
Contact service
F
Alarm
592
Special event 11
Contact service
F
Alarm
Diagnostic of process
830
Sensor temperature too
high
Reduce ambient temp. around
the sensor housing
S
Warning
831
Sensor temperature too
low
Increase ambient temp. around
the sensor housing
S
Warning