Micropilot M FMR245 with PROFIBUS PA
Operation
Hauser
41
5.4
Display and acknowledging error messages
Types of errors
Errors that occur during commissioning or measuring are displayed immediately on the local
display. If two or more system or process errors occur, the error with the highest priority is the one
shown on the display.
The measuring system distinguishes between two types of errors:
•
A (Alarm):
Instrument goes into a defined state (e.g. MAX 22 mA)
Indicated by a constant symbol.
(For a description of the codes, see Page 85)
•
W (Warning):
Instrument continue measuring, error message is displayed.
Indicated by a flashing symbol.
(For a description of the codes, see Page 85)
•
E (Alarm / Warning):
Configurable (e.g. loss of echo, level within the safety distance)
Indicated by a constant/flashing symbol.
(For a description of the codes, see Page 85)
5.4.1
Error messages
Error messages appear as four lines of plain text on the display. In addition, a unique error code is
also output. A description of the error codes is given on Page 85.
• The "
diagnostics
"
(0A)
function group can display current errors as well as the last errors that
occurred.
• If several current errors occur, use
O
or
S
to page through the error messages.
• The last occurring error can be deleted in the "
diagnostics
"
(0A)
function group
with the funktion"
clear last error
"
(0A2)
.
⇒
E HAUSER
E
+
–
Содержание Micropilot M FMR245
Страница 57: ...Micropilot M FMR245 with PROFIBUS PA Operation Endress Hauser 57 Tank linearisation MicropilotM en 307 ...
Страница 107: ...Micropilot M FMR245 with PROFIBUS PA Technical data Endress Hauser 107 ...
Страница 116: ...Micropilot M FMR245 with PROFIBUS PA Index 116 Endress Hauser ...
Страница 118: ...www endress com worldwide BA252F 00 en 05 08 71073076 CCS FM SGML 6 0 ProMoDo 71073076 ...