Commissioning
Micropilot M FMR245 with PROFIBUS PA
72
Hauser
display (008)
The
distance
measured from the reference point to the product surface and the
level
calculated
with the aid of the empty adjustment are displayed. Check whether the values correspond to the
actual level or the actual distance. The following cases can occur:
• Distance correct
−
level correct
→
continue with the next function, "
check distance
"
(051)
• Distance correct
−
level incorrect
→
Check "
empty calibr.
"
(005)
• Distance incorrect
−
level incorrect
→
continue with the next function, "
check distance
"
(051)
⇒
E HAUSER
E
+
–
⇒
⇓
After 3 s, the following message appears
E HAUSER
E
+
–
Содержание Micropilot M FMR245
Страница 57: ...Micropilot M FMR245 with PROFIBUS PA Operation Endress Hauser 57 Tank linearisation MicropilotM en 307 ...
Страница 107: ...Micropilot M FMR245 with PROFIBUS PA Technical data Endress Hauser 107 ...
Страница 116: ...Micropilot M FMR245 with PROFIBUS PA Index 116 Endress Hauser ...
Страница 118: ...www endress com worldwide BA252F 00 en 05 08 71073076 CCS FM SGML 6 0 ProMoDo 71073076 ...