Electrical connection
Memosens CFS51
24
Hauser
6.3
Post-connection check
Device health and specifications
Action
Is the outside of the sensor, assembly or cable free
from damage?
‣
Perform a visual inspection.
Electrical connection
Action
Are the mounted cables strain-relieved and not
twisted?
‣
Perform a visual inspection.
‣
Untwist the cables.
Is a sufficient length of the cable cores stripped, and
are the cores positioned in the terminal correctly?
‣
Perform a visual inspection.
‣
Pull gently to check they are seated correctly.
Are all screw terminals tightened?
‣
Tighten the screw terminals.
Are all cable entries mounted, firmly tightened and
leak-tight?
‣
Perform a visual inspection.
In the case of lateral cable entries:
‣
Point cable loops downward so that water can drip
off.
Are all cable entries mounted on the side or pointing
downwards?