background image

BA194C/07/en/12.05
51500268
valid as of:
Liquisys Software-version 2.50

Operating Instructions

Liquisys M CPM223/253

Transmitter for pH and Redox

Содержание Liquisys M CPM223

Страница 1: ...BA194C 07 en 12 05 51500268 valid as of Liquisys Software version 2 50 Operating Instructions Liquisys M CPM223 253 Transmitter for pH and Redox ...

Страница 2: ... transmitter Here you can find out how to connect digital sensors to the transmitter Æ Operation Page 28 ff Page 33 ff Page 35 ff Page 65 ff The display and operating elements are described here The operating concept is explained here The system configuration is explained here You can find information on how to calibrate the sensor on these pages Æ Maintenance Page 78 ff Page 81 ff Page 86 ff Page...

Страница 3: ... 28 5 2 Display and operating elements 28 5 2 1 Display 28 5 2 2 Operating elements 29 5 2 3 Key assignment 30 5 3 Local Operation 32 5 3 1 Automatic manual mode 32 5 3 2 Operating concept 33 5 4 System configuration 35 5 4 1 Setup 1 pH Redox 35 5 4 2 Setup 2 temperature 37 5 4 3 Current input 38 5 4 4 Current outputs 41 5 4 5 Monitoring functions 45 5 4 6 Relay contact configuration 50 5 4 7 Serv...

Страница 4: ... 9 4 Instrument specific errors 92 9 5 Spare parts 94 9 5 1 Panel mounted instrument 95 9 5 2 Field instrument 97 9 6 Return 99 9 7 Disposal 99 10 Technical data 100 10 1 Input 100 10 2 Output 100 10 3 Power supply 101 10 4 Performance characteristics 101 10 5 Environment 102 10 6 Mechanical construction 102 10 7 Documentation 102 11 Appendix 104 Index 109 ...

Страница 5: ...t have read and understood these Operating Instructions and must adhere to them Before commissioning the entire measuring point check all the connections for correctness Ensure that electrical cables and hose connections are not damaged Do not operate damaged products and secure them against unintentional commissioning Mark the damaged product as being defective Measuring point faults may only be ...

Страница 6: ...ls Warning This symbol alerts you to hazards They can cause serious damage to the instrument or to persons if ignored Caution This symbol alerts you to possible faults which could arise from incorrect operation They could cause damage to the instrument if ignored Note This symbol indicates important items of information Direct Current DC A terminal at which DC is applied or through which DC flows ...

Страница 7: ... 5 100 V AC 6 24 V AC DC ATEX II 3G EEx nAL IIC for CPM223 EEx nA L IIC T4 for CPM253 7 24 V AC CSA Gen Purp 8 24 V AC DC Output 0 1 x 20 mA pH ORP 1 2 x 20 mA pH ORP selectable 3 PROFIBUS PA 4 PROFIBUS DP 5 1 x 20 mA pH ORP HART 6 2 x 20 mA pH ORP HART selectable Additional contacts analogue input 05 not selected 10 2 x relay limit P ID timer 15 4 x relay limit P ID Chemoclean 16 4 x relay limit ...

Страница 8: ... Safety instructions for use in explosion hazardous areas XA 194C 07 a3 The delivery of the panel mounted instrument includes 1 transmitter CPM223 1 set of plug in screw terminals 2 tensioning screws 1 BNC plug solder free 1 operating instructions BA 194C 07 en versions with HART communication 1 operating instructions Field Communication with HART BA 208C 07 en versions with PROFIBUS communication...

Страница 9: ...ectrical connection section Connect the transmitter as illustrated in the Electrical connection section Commission the transmitter as explained in the Commissioning section 3 1 1 Measuring system A complete measuring systems comprises The transmitter Liquisys M CPM223 or CPM253 A pH ORP electrode with or without an integrated temperature sensor An immersible flow or retractable assembly A measurin...

Страница 10: ...ction and humidity protection The original packaging offers the best protection Also keep to the approved ambient conditions see Technical data If you have any questions please contact your supplier or your sales centre responsible 3 3 Installation conditions 3 3 1 Field instrument C07 CxM253xx 06 06 00 en 001 EPS Fig 4 Field instrument Note There is a hole in the punching for the cable entry conn...

Страница 11: ... instrument C07 CxM223xx 06 06 00 en 001 EPS Fig 6 Panel mounted instrument C07 CxM253xx 11 06 00 xx 001 eps Fig 5 View into the field housing 1 2 3 4 Removable electronics box Partition plate Terminals Fuse 1 2 3 4 90 3 54 139 5 47 96 3 78 6 0 24 mm inch 92 3 62 149 5 5 89 ...

Страница 12: ... the open air with unprotected exposure to weather conditions a weather protection cover see Accessories is required Transmitter wall mounting C07 CxM253 11 06 00 en 002 eps Fig 7 Field instrument wall mounting For wall mounting the transmitter proceed as follows 1 Drill the bores as shown in Fig 7 2 Drive the two fixing screws through the securing bores 1 from the front 3 Mount the transmitter on...

Страница 13: ... Accessories section C07 CxM253xx 11 06 00 en 003 eps Fig 8 Post mounting field device to cylindrical pipes For post mounting the transmitter proceed as follows 1 Guide the two securing screws 1 of the mounting kit through the openings of the securing plate 3 2 Screw the securing plate onto the transmitter using the four fixing screws 2 3 Secure the retainer with the field device using the clip on...

Страница 14: ... as accessories see Accessories section C07 CxM253xx 11 06 00 xx 004 eps Fig 9 Mounting field device with universal posts and weather protection cover For mounting the weather protection cover proceed as follows 1 Screw the weather protection cover with 2 screws bores 1 to the upright post bores 2 2 Secure the field device to the weather protection cover To do so use the bores 3 3 1 2 3 ...

Страница 15: ... 06 00 en 001 eps Fig 10 Securing the panel mounted instrument 1 Wall of the cabinet 2 Seal 3 Clamping screws Required installation depth 3 5 Post installation check After installation check the transmitter for damage Check whether the transmitter is protected against moisture and direct sunlight 1 2 3 3 3 92 3 62 0 5 0 02 92 3 62 0 5 0 02 90 3 54 57 2 24 33 1 30 approx 25 0 98 139 5 47 max 45 1 7...

Страница 16: ...at there is no voltage at the power cable before beginning the connection work The electrical connection of the transmitter differs depending on the device version If you are using a device without Memosens functionality please read the instructions in the Electrical connection without Memosens functionality section If you are using a device with Memosens functionality please read the instructions...

Страница 17: ...F G H I Standard sensor ISFET sensor Outer screen connection for glass electrodes Potential matching Temperature sensor Signal output 1 pH redox Signal output 2 temperature pH redox or controller Binary input 1 Hold Binary input 2 Chemoclean J K L M N O P Q Auxiliary voltage output Alarm contact position currentless Relay 1 contact position currentless Relay 2 contact position currentless Relay 3 ...

Страница 18: ...unted instrument connection sticker 1 Ground terminal for IS device version pH Redox Glas 42 43 41 54 55 56 NC SRC DRN 31 11 12 32 13 PA PM 34 33 S REF Glas IsFET IsFET IsFET NC 93 86 85 94 81 82 NC 51 52 53 57 58 59 47 48 49 pH Redox Temp opt 15V 10mA YE WH RD GN BN REL 4 SENSOR REL 3 REL 2 REL 1 NC Mains Hilfsenergie 131094 4G Digital 2 Digital 1 11 13 12 L1 N AC DC Contacts max 2A AC 250V 500VA...

Страница 19: ...nd junction boxes please refer to the Accessories section Sensor type Cable Extension Electrode without temperature sensor CPK1 VBA VBM box CYK71 cable Electrode with temperature sensor Pt 100 and TOP 68 plug in head CPK9 VBA VBM box CYK71 cable ISFET sensor with temperature sensor Pt 100 Pt 1000 and TOP 68 plug in head CPK12 VBA VBM box CYK12 cable pH individual electrode with separate reference ...

Страница 20: ...ied with the device Proceed as follows 1 Cut off end sleeves 2 and 3 of the coaxial cable Fig 15 C07 CPM223xx 04 06 00 xx 013 eps Fig 15 Cable CPK1 device connection 1 Coaxial cable 2 Inner screen BK ref 3 Inner coax pH mV 4 Strand BN PM 2 Push the cable gland 5 and the washer 6 onto the coaxial cable 3 Remove the insulation 13 mm 0 51 and screw the clamping ring 7 onto the insulation Note Parts 5...

Страница 21: ...Push the BNC connector housing 9 over the cable The inner conductor must be located at clamping surface 10 of the connector 9 Tighten the cable gland 5 10 Insert the clamp element 11 and screw in the connector cover 12 This creates a safe connection between the inner conductor and the connector pin C07 CPM223xx 04 06 00 xx 016 eps Fig 18 Mounting the pH connecting cable in the BNC elbow plug 5 9 1...

Страница 22: ...ential matching If you want asymmetrical measurement you have to change the configuration in the A2 field If the asymmetrical software setting was selected for a symmetrical connection the operating time of the reference electrode is reduced Caution In the case of a symmetrical connection the potential matching pin must be connected and always immersed in the medium Advantages of symmetrical vs as...

Страница 23: ...ransmitter C07 CPM2x3xx 04 06 00 xx 002 esp Fig 21 Transmitter electrical connection with Memosens technology A B C D E F G Screen Sensor Signal output 1 pH redox Signal output 2 temperature pH redox or controller Binary input 1 Hold Binary input 2 Chemoclean Auxiliary voltage output H I J K L M N Alarm contact position currentless Relay 1 contact position currentless Relay 2 contact position curr...

Страница 24: ...Memosens functionality Caution Terminals marked NC may not be switched Unmarked terminals may not be switched Note Please label the sensor terminal block with the TU sticker provided Do not use the pH sticker 42 43 41 54 55 56 NC 96 88 87 31 NC 97 NC 32 NC 34 33 S B NC 93 86 85 94 81 82 NC 51 52 53 57 58 59 47 48 49 pH Temp opt 15V 10mA GN BN YE WH A REL 4 SENSOR REL 3 REL 2 REL 1 136771 0003 4A D...

Страница 25: ... refer to the Accessories section To plug the cable coupling onto the sensor plug in head proceed as follows YE GN BN WH 87 88 Com A 96 Com B 97 1 C07 MemoSnxx 11 05 00 xx 002 eps Fig 25 Handling of sensor plug in head and cable coupling 1 Rotate the lower part of the coupling in such a way that the two pairs of keys pos 1 2 are located above each other 2 Plug the coupling onto the plug in head so...

Страница 26: ...r of the device see connection sticker Example of connecting pH electrode The following diagram shows the connection of a pH electrode with Memosens functionality C07 CPM2x3xx 04 06 00 xx 012 eps Fig 26 Connecting CPS11D with CYK10 Signal transmission between the Memosens electrode and the coupling of the CYK10 cable takes place contact free and via completely encapsulated coils This offers the fo...

Страница 27: ...larm condition Error message present alarm LED red or Device defective or voltage free alarm LED off Relay de energised Contact 41 42 closed Device status and specifications Remarks Are the transmitter or the cable externally damaged Visual inspection Electrical connection Remarks Are the installed cables strain relieved No loops and cross overs in the cable run Are the signal cables correctly con...

Страница 28: ...lease read the relevant sections in the additional Operating Instructions PROFIBUS PA DP field communication for Liquisys M CXM223 253 BA 209C 07 en HART field communication for Liquisys M CXM223 253 BA 208C 07 en The following section only explains operation via the keys 5 2 Display and operating elements 5 2 1 Display LED display Indicates the current operating mode Auto green LED or Manual yell...

Страница 29: ... display 1 2 3 4 5 6 7 Indicator for measuring mode normal operation Indicator for calibration mode Indicator for setup mode configuration Indicator for Hold mode current outputs remain at last current state Indicator for receipt of a message for devices with communication Indicator of working status of relays 3 4 d inactive c active Function code display 8 9 10 11 12 13 14 In measuring mode measu...

Страница 30: ...he System configuration section Configuration of parameters and numerical values Operation of the relay in manual mode In the measuring mode you get the following sequence of functions by repeatedly pressing the PLUS key 1 Temperature display in F 2 Temperature display hidden 3 Measured value display in mV 4 Current input signal in 5 Current input signal in mA 6 Return to basic settings In the mea...

Страница 31: ... the end of calibration if calibrating If you press the PLUS and MINUS key again you return to the measuring mode Locking the keyboard Press the PLUS and ENTER key for at least 3 s to lock the keyboard against any unauthorised data entry All the settings can continue to be read The code prompt displays the code 9999 Unlocking the keyboard Press the CAL and MINUS key for at least 3 s to unlock the ...

Страница 32: ... manual mode 1 The transmitter is in Automatic mode The top LED beside the AUTO key is lit 2 Press the AUTO key The bottom LED beside the AUTO key lights up 3 To enable the manual mode enter the code 22 via the PLUS and MINUS keys 4 Select the relay or the function You can use the REL key to switch between the relays The relay selected and the switch status ON OFF is displayed on the second line o...

Страница 33: ... access to Calibration and Offset menu Key ENTER Code 22 access to the menus for the parameters which make configuration and user specific settings possible Keys PLUS ENTER locks the keyboard Keys CAL MINUS unlocks the keyboard Key CAL or ENTER any code access to read mode i e all the settings can be read but not modified The device continues measuring in the read mode It does not shift to the Hol...

Страница 34: ... C07 CxM2x3xx 19 06 00 xx 002 eps Fig 31 Diagram of the menu structure 1 Functions parameters selected numbers entered 2 Function groups scroll backwards and forwards with the PLUS and MINUS keys 3 Switch from function to function with the ENTER key Hold function freezing of the outputs In both the setup mode and during calibration the current output can be frozen it constantly retains its current...

Страница 35: ... are automatically reset to the factory settings A2 Select connection mode Sym symmetrical Asym asymmetrical Detailed information on symmetrical or asymmetrical connections can be found in the Sensor connection section A3 Enter measured value damping 1 1 60 Measured value damping causes averaging over the number of individual measured values entered This is used for example to stabilise the displa...

Страница 36: ... contact galvanically isolated signal transmission only simple asymmetrical connection necessary A3 Enter measured value damping 1 1 60 Measured value damping causes averaging over the number of individual measured values entered This is used for example to stabilise the display if the measurement is unstable There is no damping if 1 is entered A4 Sensor Glass No edit option Note Glass electrodes ...

Страница 37: ... process temperature which is to be used for compensation B2 Enter process temperature 25 0 C 50 0 150 0 C Only if A1 pH and B1 MTC You can edit the displayed value The value entered can only be in C B3 Select the type of temperature compensation for the calibration ATC MTC For B1 ATC edit possible For B1 MTC only display B3 MTC return to B A separate temperature sensor must also be immersed in th...

Страница 38: ...ing of an alarm condition in the main stream flow too low or has completely stopped and triggers dosing switch off even if the medium flow is retained due to the method of installation Flow in main stream Current signal in mA Current input signal in Current input lower range limit Flowmeter lower setting value 4 0 Current input upper range limit Flowmeter upper setting value 20 100 C07 CPM2x3xx 05...

Страница 39: ... below example with factory setting C07 CPM2x3xx 05 06 00 de 003 eps Fig 34 Multiplying feedforward control Y Gain Kinfl X Current input signal C07 CPM2x3xx 05 06 00 xx 002 eps Fig 33 Sample arrangement for feedforward control of the flow in the main stream to the PID controller 1 2 3 4 Measuring water extraction point Static mixer Injection points Flowmeter 5 6 7 8 Alkali Acid Liquisys M CPM253 C...

Страница 40: ... Z4 Enter the switch off limit value for the current input 50 0 100 0 100 corresponds to 4 20 mA at the current input Observe measured value assignment to the current output of the flowmeter Z5 Enter the switch off direction for the current input Low High The controller is switched off if the value entered in Z4 is undershot or overshot Z6 Select feedforward control to PID controller Off Lin linea...

Страница 41: ...rrent output C07 CPM2x3xx 05 06 00 xx 004 eps Fig 35 User defined current output characteristic example The current output characteristic must be extremely monotonic increasing or extremely monotonic decreasing The distance per mA between two table value pairs must be greater than pH 0 03 Redox 5 mV Temperature 0 25 C The values for the sample characteristic Fig 35 are entered in the following tab...

Страница 42: ...selected for every output O2 Select measured variable for 2nd current output C pH mV Contr R237 R 266 curr current output 2 can only be selected if O2 Contr is selected relay board required O3 O3 1 Enter or output linear characteristic Lin linear 1 Sim simulation 2 Tab table 3 The characteristic can have a positive or negative slope for the measured value output In the case of actuating variable o...

Страница 43: ... O3 3 O321 Enter simulation value Current value 0 00 22 00 mA Entering a current value results in this value being directly output at the current output O3 3 Enter current output table only for Plus Package Lin linear 1 Sim simulation 2 Tab table 3 Values can also be added or altered at a later stage The values entered are automatically sorted by increasing current value For further characteristic...

Страница 44: ...fied by user Return to O333 until all values are entered O336 Message as to whether table status is OK yes no Back to O3 If status no correct table all settings made up until now are retained or back to measuring mode table is deleted Coding Field Setting range Factory settings bold Display Info O335 mA mA value 0 00 O336 Status ok yes ...

Страница 45: ...F2 Select time unit s min F3 Enter alarm delay 0 s min 0 2000 s min Depending on the option selected in F2 the alarm delay is entered in s or min F4 Select error current 22 mA 2 4 mA This selection must be made even if all error reporting is switched off in F5 Caution If 0 20 mA was selected in O311 2 4 mA may not be used F5 Select error 1 1 255 Here you can select all the errors which should trig...

Страница 46: ...impedance is undershot approx 500 kΩ This function cannot be selected for antimony and ISFET electrodes A4 Monitoring of impedance of the reference electrode alarm if set threshold value overshot This function can only be selected for symmetrically high resistance measurement Monitoring of leak current for ISFET sensors pre alert E168 at ILEAK 200 nA error E008 at ILEAK 400 nA F7 Set error current...

Страница 47: ...ller activity with the function CC A malfunction of the controller is detected and reported thanks to freely adjustable monitoring times E154 E157 C07 CxM2x3xx 05 06 00 xx 001 eps Fig 37 PCS alarm live check A Constant measuring signal alarm triggered after PCS alarm time has elapsed Note The electrode must be symmetrically connected with PML to monitor the reference Any PCS alarm pending is autom...

Страница 48: ... the reference electrode increases with the degree of contamination Not for Memosens P4 Leak current display for ISFET sensor 0 0 9 9 µA Only if A4 ISFET and P1 on Display only Leak currents 0 4 µA indicate damage to the ISFET sensor P5 Select alarm threshold monitoring Off Low High LoHi low high Low High LoHi Alarm possible with or without controller switch off xxxx without controller switch off ...

Страница 49: ...s monitoring PCS alarm Off AC CC AC CC AC CC ACCC AC sensor activity monitoring CC controller monitoring Alarm possible with or without simultaneous controller switch off xxxx without controller switch off xxxx with controller switch off P10 Enter maximum permissible duration for alarm threshold undershoot 60 min 0 2000 min Only for P9 CC or AC CC P11 Enter maximum permissible duration for alarm t...

Страница 50: ...mperature measurement Please refer to Abb 38 for a clear illustration of the relay contact states When the measured values increase maximum function the relay contact is closed as of t2 after the switch on point t1 has been overshot and the pick up delay has elapsed t2 t1 The alarm contact switches if the alarm threshold t3 is reached and the alarm delay t4 t3 has also elapsed When the measured va...

Страница 51: ...PID controller The following configuration options are available for a PID controller Change control gain Kp P influence Set integral action time Tn I influence Set derivative action time Tv D influence Basic load dosing Basic The basic load dosing field R231 is used to set a constant dosage field R2311 PID controlling plus basic load dosing If you select this function PID Basic in field R231 the ...

Страница 52: ...er the switching frequency of the contact affected The maximum switching frequency 1 T can be set between 60 and 180 min 1 The on time tON is constant It depends on the set maximum frequency and is approx 0 5 s for 60 min 1 and approx 170 ms for 180 min 1 Outputs with pulse frequency modulation are used to activate directly controlled solenoid dosing pumps C07 CxM2x3xx 05 06 00 xx 004 eps Fig 40 O...

Страница 53: ...n Timer for cleaning function This function includes a simple cleaning option You can set the time interval after which cleaning should start So you can only select a constant interval sequence Other cleaning functions are available for selection in conjunction with the Chemoclean function version with four contacts see Chemoclean function section Note Timer and Chemoclean do not work independentl...

Страница 54: ...be equipped with a designated relay board see product structure or chapter accessories Timer and Chemoclean do not work independently of one another While one of the two functions is active the other cannot be started For the Chemoclean function the relays 3 water and 4 cleaner are used If the cleaning is prematurely aborted a post rinse time always follows If the setting is Economy cleaning only ...

Страница 55: ...d as desired in the X direction via set point 1 and 2 Note Neutralisation control is only possible with relays 1 and 2 C07 CxM2x3xx 05 06 00 xx 009 eps Fig 45 Control characteristic of a proportional neutralisation controller 1 2 3 Neutral zone Set point 1 Set point 4 5 6 Set point 2 Control contact 1 for alkali Control contact 2 for acid Y X 100 50 0 1 2 3 4 5 6 Coding Field Setting range Factory...

Страница 56: ...the alarm threshold is undershot overshot this triggers an alarm with the error message and error current at the transmitter note alarm delay in field F3 If defined as a Min contact the alarm threshold must be switch off point R217 Display status for limit contactor MAX MIN Display only R2 2 Configure limit contactor for temperature measurement LC PV limit contactor pH redox 1 LC C limit contactor...

Страница 57: ...IN Display only R2 3 Configure P ID controller LC PV limit contactor pH redox 1 LC C limit contactor T 2 PID controller 3 Timer 4 Clean Chemoclean 5 Neutra controller By confirming with ENTER another relay function already switched on is switched off and its settings are reset to the factory settings R231 Switch function of R2 3 off or on Off On Basic PID B On PID controller Basic basic load dosin...

Страница 58: ...ength is selected in R237 If pulse frequency is selected R238 is skipped and entries continue with R239 R239 Enter maximum pulse frequency of the adjuster 120 min 1 60 180 min 1 This field only appears if pulse frequency is selected in R237 If pulse length is selected R239 is skipped and entries continue with R2310 R2310 Enter minimum switch on time tON 0 3 s 0 1 5 0 s This field only appears if p...

Страница 59: ...g with Chemoclean for version with four contacts and contacts 3 and 4 assigned LC PV limit contactor pH redox 1 LC C limit contactor T 2 PID controller 3 Timer 4 Clean Chemoclean 5 Neutra controller 6 See Chemoclean function section By confirming with ENTER another relay function already switched on is switched off and its settings are reset to the factory settings R251 Switch function of R2 5 off...

Страница 60: ...ut with water only until the next cleaning session with cleaner takes place R2 6 Configure neutralisation controller LC PV limit contactor pH redox 1 LC C limit contactor T 2 PID controller 3 Timer 4 Clean Chemoclean 5 Neutra controller 6 Only for A1 pH If neutra controller is selected for Rel1 only neutra controller is offered for Rel2 By confirming with ENTER another relay function already switc...

Страница 61: ...val 10 0 s 0 5 999 9 s This field only appears if pulse length is selected in R266 If pulse frequency is selected R267 is skipped and entries continue with R268 R268 Enter maximum pulse frequency of the adjuster 120 min 1 60 180 min 1 This field only appears if pulse frequency is selected in R266 If pulse length is selected R268 is skipped and entries continue with R269 R269 Enter minimum switch o...

Страница 62: ...er Hold dwell period 10 s 0 999 s S5 Enter SW upgrade release code Plus Package 0000 0000 9999 If an incorrect code is entered you are taken back to the measurement menu The number is edited with the PLUS or MINUS key and confirmed with the ENTER key 1 is displayed if the code is active S6 Enter SW upgrade release code Chemoclean 0000 0000 9999 If an incorrect code is entered you are taken back to...

Страница 63: ...ue is fluctuating or sporadic SCS errors are reported Coding Field Setting range Factory settings bold Display Info S9 S Default no S10 Test no S11 mV uncompRef 25 S12 Hz AC Freq 50 Coding Field Setting range Factory settings bold Display Note E E H SERVICE function group Information on the device version E1 Select module Contr controller 1 Trans transmitter 2 Main power unit 3 Rel relay module 4 ...

Страница 64: ...ting range Factory settings bold Display Note E114 Modul ID LSG Coding Field Setting range Factory settings bold Display Info I INTERFACE function group Communication settings only for device version HART or PROFIBUS I1 Enter bus address Address HART 0 15 or PROFIBUS 0 126 Each address may only be given once in a network If a device address 0 is selected the current output is automatically set to ...

Страница 65: ...r the calibration is accepted If the slope or zero point are outside the ranges given in C16 and C17 error 32 becomes active for slope or error 33 becomes active for zero point The electrode must then be checked and replaced if necessary If precalibrated digital sensors Memosens functionality are connected the calibration data are automatically transmitted to the transmitter Things to note element...

Страница 66: ...constant value over 10 s If the value does not stabilise within 5 min error 44 is set and the calibration is aborted In the case of symmetrical measuring operation the potential matching pin must also be immersed in the buffer C13 Calibration is performed Stability check The value is accepted in the event of stability pH 0 05 for more than 10 s C14 Enter pH value of the second buffer solution Buff...

Страница 67: ...he mV value belonging to the redox buffer used Current measured value 1500 1500 mV In the case of symmetrical measuring operation the potential matching pin must also be immersed in the buffer C22 Calibration is performed mV value Stability check The value is accepted in the event of stability 1 mV for more than 10 s C23 Zero point is displayed 100 100 mV C24 Calibration status is displayed o k E ...

Страница 68: ...onfirmed with the CAL key see calibration pH C32 Calibration is performed mV value is displayed Accepted in the event of stability 5 mV for more than 10 s C33 Determine 20 value of the non toxic sample 20 0 100 The procedure for C31 is repeated with the non toxic sample to calibrate value 2 C34 Calibration is performed mV value is displayed Stability check The value is accepted in the event of sta...

Страница 69: ...pe Glass 59 16 mV pH 38 00 65 00 mV pH Antimony 59 16 mV pH 25 00 65 00 mV pH ISFET 59 16 mV pH 38 00 65 00 mV pH For A4 ISFET enter the slope from the quality certificate N3 Enter zero point Glass 7 00 pH 5 00 9 00 pH Antimony 1 00 pH 1 00 3 00 pH ISFET 0 mV 500 500 mV For A4 ISFET enter the voltage UIS from the quality certificate N4 Calibration status is displayed o k E xxx N5 Store calibration...

Страница 70: ... X A B Coding Field Setting range Factory settings bold Display Info V OFFSET function group for pH or redox Depending on the operating mode selected either pH or redox appears i e no direct selection can be made V1 Enter desired measured value Current measured value pH 2 00 16 00 1500 1500 mV 0 0 100 0 You can edit the display The entry can differ from the actual value by max 2 0 pH 120 mV 50 V2 ...

Страница 71: ...de on the accuracy 7 Check the measured value transfer to the process control system or the signal processing unit 6 2 Things to note when commissioning ISFET sensors Switch on behaviour A control circuit is created when the measuring system is switched on During this time approx 5 8 minutes the measured value adjusts to the real value This settling behaviour occurs every time the liquid film betw...

Страница 72: ...easuring system can return precise measurement data Then perform the first configuration in accordance with the instructions in the Quick start up section The values set by the user are kept even in the event of a power failure The following function groups are available in the transmitter the groups only available in the Plus Package are marked accordingly in the functional description Setup mode...

Страница 73: ...edox 1500 mV or 0 Limit value for controller 2 pH 16 redox 1500 mV or 100 Hold Active during configuration and calibration Contact 1 4 Limit contactor pH function off Current outputs 1 and 2 4 20 mA Current output 1 measured value for 4 mA signal current pH 2 Current output 1 measured value for 20 mA signal current pH 12 Current output 2 temperature value for 4 mA signal current 0 0 C 32 F Current...

Страница 74: ...ake your settings for Setup 1 9 In A1 select the desired mode of operation e g pH Press F to confirm pH ORP redox mV ORP redox 10 In A2 select the type of connection for your sensor See the Sensor connection section for this Press F to confirm sym symmetrical asym asymmetrical 11 In A3 enter the damping factor Measured value damping averages the individual measured values and serves to stabilise t...

Страница 75: ...n e g ATC for automatic temperature compensation Press F to confirm ATC MTC 18 The current temperature is displayed in B4 If necessary calibrate the temperature sensor to an external measurement Press F to confirm Actual value displayed and entered 50 0 150 0 C 19 The difference between the measured and entered temperature is displayed Press F The display returns to the initial display of the Setu...

Страница 76: ... accuracy and reliability are also guaranteed after repair Note If you have any queries please contact your local Sales Office 7 1 Maintenance transmitter 7 1 1 Dismantling of panel mounted instrument Caution Please note the effects on the process if the device is taken out of service Note Please refer to the diagram in Section 9 5 for the item numbers 1 Disconnect the terminal block item 422 b fr...

Страница 77: ...ion jack Using a fine side cutting pliers nip off the heads of the synthetic distance sleeves Then remove the module from above To assemble carefully push the modules into the trolley tracks of the electronics box and let them engage in the side box noses Note Incorrect mounting is not possible Modules inserted in the electronics box incorrectly are not operable since the ribbon cables cannot be c...

Страница 78: ...d An incorrect serial number can only be corrected at the factory Press ENTER to confirm the serial number or cancel the entry to enter the number again 7 If available enter the release codes for the Plus Package and or Chemoclean in the Service menu 8 Check the Plus Package release e g by opening the function group CHECK Code P or the Chemoclean function 9 Make the customer device settings again ...

Страница 79: ...installation or installation with the plug in head pointing downwards C07 CPM2x3xx 05 06 00 xx 006 eps Fig 48 Permitted angle of installation for glass electrodes Redox electrodes Carefully clean the metal pins or surfaces mechanically Note After mechanical cleaning the redox sensor can require several hours conditioning time For this reason check the calibration after a day 7 2 3 Maintaining digi...

Страница 80: ...le Caution If there is moisture in the measuring cable the cable must be replaced A shunt in the cable of 20 MΩ can no longer be measured with normal multimeters but is damaging for the pH measurement A reliable test can be carried out with a usual commercial insulation meter Make sure to disconnect the pH measuring cable from the sensor and device If you are using a junction box you check check t...

Страница 81: ...nical Information TI374 C07 en Orbipore CPS91 CPS91D pH electrode with open aperture for media with high dirt load Memosens technology as option Ordering acc to product structure see Technical Information TI375C 07 en Tophit CPS471 Sterilisable and autoclavable ISFET sensor for food and pharmaceuticals process technology water treatment and biotechnology Ordering acc to product structure see Techn...

Страница 82: ...rminals protection class IP 65 material polycarbonate order no 50005276 8 3 Mounting accessories Weather protection cover CYY101 for mounting of field housing for outdoor installation material stainless steel 1 4031 order no CYY101 A C07 CYY101xx 00 06 00 en 001 eps Fig 49 Weather protection cover for field instrument Certificates A Standard non Ex Cable length 03 Cable length 3 m 9 84 ft 05 Cable...

Страница 83: ...102xx 00 06 00 en 002 eps Fig 50 Square post CYY102 Kit for mounting of field housing on horizontal or vertical pipes Ø max 60 mm 2 36 order no 50086842 C07 CxM2x3xx 00 06 00 en 001 eps Fig 51 Pipe mounting kit 1495 58 86 60 x 60 2 36 x 2 36 120 4 72 150 5 91 mm inch 70 2 76 90 3 54 Ø 6 0 24 Ø 8 0 31 70 2 76 90 3 54 M6 60 2 36 90 3 54 70 2 76 Ø 8 0 31 30 1 18 Ø max 60 2 36 mm inch ...

Страница 84: ...Order as per product structure see Technical Information TI 344C 07 en Cleanfit P CPA474 Process retractable assembly made of plastic with ball valve shut off for particularly safe separation of the process medium from the environment Order as per product structure see Technical Information TI 345C 07 en Dipfit W CPA111 Immersion and installation assembly made of plastic for open and closed tanks ...

Страница 85: ...chnical buffer solutions for ORP electrodes 220 mV pH 7 0 100 ml 0 026 US gal order no CPY3 0 468 mV pH 0 1 100 ml 0 026 US gal order no CPY3 1 KCl electrolyte solutions for liquid filled electrodes 3 0 mol T 10 100 C 14 212 F 100 ml 3 oz order no CPY4 1 3 0 mol T 10 100 C 14 212 F 1000 ml 30 oz order no CPY4 2 1 5 mol T 30 100 C 22 266 F 100 ml 3 oz order no CPY4 3 1 5 mol T 30 100 C 22 266 F 100...

Страница 86: ...ibration data invalid no user data invalid EEPROM error instrument software not suitable to hardware controller Yes No 1 E003 Download error Invalid configuration Repeat download check optoscope Yes No No E004 Instrument software version not compatible with module hardware version Load software compatible with hardware Load measurement parameter specific device software Yes No No E007 Transmitter ...

Страница 87: ...t output range 1 exceeded Yes No No E065 Below current output range 2 Yes No No E066 Current output range 2 exceeded Yes No No E067 Set point exceeded controller 1 Check configuration Yes No No E068 Set point exceeded controller 2 Yes No No E069 Set point exceeded controller 3 Yes No No E070 Set point exceeded controller 4 Yes No No E080 Current output 1 range too small Increase range in Current o...

Страница 88: ...replace as soon as possible No No No E171 Flow in main stream too low or zero Restore flow Yes No No E172 Switch off limit for current input exceeded Check process variables at sending measuring instrument Change range assignment if necessary Yes No No E173 Current input 4 mA Check process variables at sending measuring instrument Yes No No E174 Current input 20 mA Check process variables at sendi...

Страница 89: ...sensor Measuring chain slope not adjustable no slope Hair line crack in the glass membrane Renew sensor pH sensor No high impedance connection moisture dirt Check cable plug connector and junction boxes pH simulator insulation see Checking the connecting lines and boxes section Permanent incorrect measured value Sensor not immersing or protection cap not removed Check installation position remove ...

Страница 90: ...e Switch off simulation See field O2 Impermissible processor system operating status Switch device off and then on again EMC problem if it persists check the installation Incorrect current output signal Incorrect current assignment Check current assignment 0 20 mA or 4 20 mA Field O211 Overall burden in the current loop too high 500 Ω Disconnect output and measure directly at the device mA metre f...

Страница 91: ...tion No PA DP central module Check using nameplate PA xxx3xx DP xxx4xx Retrofit to LSCP module see Spare parts section Incorrect device software version without PROFIBUS For further information see BA 209C 07 en PROFIBUS PA DP field communication with Liquisys M CxM223 253 With Commuwin CW II CW II version and device software version incompatible No or incorrect DD DLL Baudrate for segment coupler...

Страница 92: ...ed Device or module in device not correctly mounted CPM223 reinstall insert CPM253 remount display module Perform with the aid of the installation drawings in the Spare parts section Operating system in unpermitted mode Switch device off and then on again Poss EMC problem if this persists check the installation or have checked by Endress Hauser Service Device gets hot Voltage wrong too high Compar...

Страница 93: ...ial number 0000 when they leave the factory The S package or Chemoclean are not enabled on leaving the factory In the case of LSCH LSCP with SNR 0000 a device serial number can be entered once in fields E114 to E116 Then enter the release code for the S package and or Chemoclean For a detailed description see Replacing central module section No HART or PROFIBUS PA DP interface function Incorrect c...

Страница 94: ...s kits To be on the safe side you should always specify the following data with your spare part orders Instrument order code order code Serial number serial no Software version where available Refer to the nameplate for the order code and serial number The software version is displayed in the instrument sofware see chapter Instrument configuration if the instrument processor system is functional ...

Страница 95: ...ounted instrument You can take the spare parts and the corresponding order number from the following section using the item numbers C07 CxM223xx 09 06 06 xx 001 eps Fig 52 Exploded drawing of panel mounted instrument Pos 10 Netzteil 100 230 VAC 230 V 100 V 115 V Sicherung Pos 30 Steckbrücke A 10 30 B 422a 422b 400 340 435 235 230 10 20 40 330 290 300 430 310 320 ...

Страница 96: ...ng plate 51501070 290 Relay module LSR1 2 2 relays 51500320 290 Relay module LSR2 2i 2 relays current input 4 20 mA 51504304 300 Relay module LSR1 4 4 relays 51500321 300 Relay module LSR2 4i 4 relays current input 4 20 mA 51504305 310 Side panel Kit with 10 parts 51502124 310 320 340 400 Housing mechanical parts Rear plate side panel end frame special screw 51501076 330 400 Housing module Housing...

Страница 97: ...device You can take the spare parts and the corresponding order number from the following section C07 CxM253xx 09 06 06 xx 001 eps Fig 53 Field device exploded drawing 100 V Pos 10 Netzteil 100 230 V AC 230 V 115 V Sicherung Pos 30 Steckbrücke 30 10 C D 410 460b 460a 420 40 310 320 230 10 20 340 450c 470 480 290 300 400 430 450a 330 370 A B 999 450b 490 ...

Страница 98: ... LSR2 2i 2 relays current input 4 20 mA 51504304 300 Relay module LSR1 4 4 relays 51500321 300 Relay module LSR2 4i 4 relays current input 4 20 mA 51504305 310 320 Ribbon cable lines 2 ribbon cable lines 51501074 340 330 450 Inner housing fittings Docking module empty electronics box small parts 51501073 450a 450c Torx screws K4x10 Part of inner housing fittings 450b Torx screw for central module ...

Страница 99: ...lease enclose the completed Declaration of contamination copy the second last page of these Operating Instructions with the packaging and the transportation documents No repair without completed Declaration of contamination 9 7 Disposal The device contains electronic components and must therefore be disposed of in accordance with regulations on the disposal of electronic waste Please observe local...

Страница 100: ...Load Max 500 Ω depending on operating voltage Transmission range pH Adjustable min 1 pH Redox Absolute Relative Adjustable min 50 mV Fixed 0 100 Temperature Adjustable 10 100 of end of measuring range Continuous controller 0 100 Y signal Signal resolution Max 700 digits mA Minimum spread 0 4 20 mA signal 10 of the measuring range span Insulation voltage Max 350 Veff 500 V DC Auxiliary voltage outp...

Страница 101: ... semi delay 250 V 3 15 A Reference temperature 25 C 77 F Measured value resolution pH 0 01 pH Redox 1 mV 0 1 Temperature 0 1 C Maximum measured error1 1 In accordance with IEC 746 1 for nominal operating conditions Display pH Redox Temperature Max 0 5 of scope of measuring range Max 0 5 of scope of measuring range Max 1 0 of scope of measuring range Signal output pH Redox Temperature Max 0 75 of s...

Страница 102: ...10 95 not condensating Design dimensions Panel mounted instrument L x W x D 96 x 96 x 145 mm 3 78 x 3 78 x 5 71 Installation depth approx 165 mm 6 50 Field device L x W x D 247 x 170 x 115 mm 9 72 x 1 70 x 4 53 Weight Panel mounted instrument Max 0 7 kg 1 5 lb Field device Max 2 3 kg 5 1 lb Material Panel mounted instrument housing Polycarbonate Field housing ABS PC Fr Front membrane Polyester UV ...

Страница 103: ...Liquisys M CPM223 253 Technical data Endress Hauser 103 ...

Страница 104: ...easured value C21 Function group SETUP 1 A Select operating mode ORP mV ORP pH A1 Select connection type asym asymmetrical sym symmetrical A2 Enter measured value damping no damping 1 60 1 A3 Select sensor Glass E 7 0 Antim Antimon ISFET 0 Select temperature sensor Pt 100 NTC 30K Pt 1k A4 A5 Function group SETUP 2 B Select temperature compensation for the process pH Redox on MTC ATC off B1 Enter M...

Страница 105: ...s displayed o k E C18 Field for customer settings Automatic start of cleaning function yes not always displayed see error messages no F8 Activate error current for previously set error yes no F7 F9 Select next error or return to menu next error R next Enter y value current value 0 00 20 00 mA 0 00 mA O335 Table status o k no yes O336 Enter x value measured valued 2 00 16 00 pH 1500 1500 mV 0 0 100...

Страница 106: ...figuration Neutr neutralisation controller with Rel1 and Rel2 and A1 pH only Clean Chemoclean with Rel3 only Timer PID controller LC C Limit contactor T LC PV Limit contactor pH Redox Function group SERVICE S Select language GER ITA FRA ESP NEL ENG S1 Hold configuration s c during setup and calibration CAL during calibration Setup during setup none no hold S2 Software version SW version E111 Hardw...

Страница 107: ...Tv 0 0 no D componentl 0 0 999 9 min 0 0 min R265 Enter minimum ON time t 0 1 5 0 s on 0 3 s R269 Enter maximum pulse frequency 60 180 1 min 120 1 min R268 Select freq pulse frequency curr current output 2 len pulse length R266 Enter pulse interval 0 5 999 9 s 10 0 s R267 Enter post rinse time 0 999 s 20 s R255 Set number of repeat cycles 0 5 0 R256 Set interval between two cleaning cycles pause t...

Страница 108: ...Appendix Liquisys M CPM223 253 108 Endress Hauser ...

Страница 109: ...tion With Memosens 23 Without Memosens 17 Electrical icons 6 EMC 6 Environment 102 F Factory settings 73 Field device connection 18 24 Freezing of outputs 34 H Hold function 34 62 I Icons Electrical 6 Safety symbols 6 Incoming acceptance 10 Input 100 Installation 5 9 10 12 Instrument specific errors 92 Interfaces 64 K Key assignment 30 L Limit contactor 50 Liquid KCl supply 80 M Maintenance 76 Ass...

Страница 110: ...on 50 Replacement of controller 77 Return 6 99 S Safety icons 6 Scope of delivery 8 SCS Operating voltage 49 Service 62 Setup 1 35 Setup 2 37 Spare parts 94 storage 10 Switching on 72 Symbols 6 System configuration 35 System error messages 86 T Technical data 100 102 Timer 53 Cleaning function 53 transport 10 Trouble shooting 86 Instructions 86 Instrument specific errors 92 Process specific errors...

Страница 111: ...Liquisys M CPM223 253 Endress Hauser 111 ...

Страница 112: ...www endress com worldwide BA194C 07 en 12 05 51500268 Printed in Germany FM SGML 6 0 DT 51500268 ...