Operation
Liquisys M CPM223/253
68
Hauser
For the calibration, a sample of the medium is placed in two containers. The
contents of the first container is detoxified. The contents of the second
container remains unchanged. A relative value of 80 % is set with the "toxic"
sample. A relative value of 20 % is set with the "non-toxic" sample.
Default values:
0 % = -1000 mV
100 % = +1000 mV
The calibration range is ±1500 mV, the
minimum difference should be 60 mV.
C31
Determine 80% value
of the "toxic" sample
80%
0 ... 100%
Start the calibration of the "toxic" sample by
pressing the CAL key. The value is accepted
provided it is stable or confirmed with the
CAL key (see calibration pH).
C32
Calibration is
performed
mV value is displayed
Accepted in the event of stability
≤
±5 mV
for more than 10 s.
C33
Determine 20% value
of the "non-toxic"
sample
20%
0 ... 100%
The procedure for C31 is repeated with the
"non-toxic" sample to calibrate value 2.
C34
Calibration is
performed
mV value is displayed
Stability check:
The value is accepted in the event of stability
≤
±5 mV for more than 10 s.
C35
Calibration status is
displayed
o.k.
E xxx
C36
Store calibration
result?
yes
no
new
If C35 = E xxx, then only No or
New
.
If New, return to C.
If Yes/No, return to "Measurement".
The electrode can now be reinstalled in the process.
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
C31
%
Sample 1
80
C32
mV
CAL 80%
200
C33
%
Sample 2
20
C34
mV
CAL 20%
50
C35
Status
o.k.
C36
Store
yes