Trouble-shooting
Liquisys M CPM223/253
88
Hauser
E154
Below lower alarm threshold for
period exceeding alarm delay
Perform manual comparison measurement if
necessary. Service sensor and recalibrate.
Yes
No
No
E155
Above upper alarm threshold for
period exceeding alarm delay
Yes
No
No
E156
Actual value undershoots alarm
threshold for longer than the set
permissible maximum period
Yes
No
No
E157
Actual value exceeds alarm threshold
for longer than the set permissible
maximum period
Yes
No
No
E162
Dosage stop
Check settings in the CURRENT INPUT and
CHECK function groups.
Yes
No
No
E164
Dynamic range of pH converter
exceeded
Check cable and sensor.
Yes
No
—
E166
Dynamic range of reference
converter exceeded
Check cable and sensor.
Yes
No
—
E168
Warning: ISFET leakage
current > 200 nA
Check ISFET for abrasion and air-tightness, replace
as soon as possible.
No
No
No
E171
Flow in main stream too low or zero
Restore flow.
Yes
No
No
E172
Switch-off limit for current input
exceeded
Check process variables at sending measuring
instrument. Change range assignment if necessary.
Yes
No
No
E173
Current input < 4 mA
Check process variables at sending measuring
instrument.
Yes
No
No
E174
Current input > 20 mA
Check process variables at sending measuring
instrument. Change range assignment if necessary.
Yes
No
No
E175
SCS glass warning
Check electrode for glass breakage or hair-line
cracks;
Check medium temperature.
Measurement can continue until the error occurs.
No
No
No
E177
SCS reference electrode warning
Check electrode for contamination and damage;
clean electrode; measuring can continue until the
error occurs.
No
No
No
1)
If this error occurs, there is no possibility of starting a cleaning session (field F8 not applicable with this error).
Error
no.
Display
Tests and/or remedial measures
Alarm contact
Error current
Autom. cleaning
start
Facty
User
Facty
User
Facty
User
Содержание Liquisys M CPM223
Страница 103: ...Liquisys M CPM223 253 Technical data Endress Hauser 103 ...
Страница 108: ...Appendix Liquisys M CPM223 253 108 Endress Hauser ...
Страница 111: ...Liquisys M CPM223 253 Endress Hauser 111 ...
Страница 112: ...www endress com worldwide BA194C 07 en 12 05 51500268 Printed in Germany FM SGML 6 0 DT 51500268 ...