Problem
Possible cause
Remedy
Equipment needed, spare parts
S package functions not
available
– S package not enabled (enable
with code that depends on serial
number and is received from
E+H with order of S package)
– S package upgrade:
code received from E+H
⇒
enter
– Following replacement of
defective LSCH/LSCP module:
first enter instrument serial
number (s. nameplate) manually,
then enter code
Detailed description see chpt. 8.5.5
No HART
communication
– No central HART module
Verify by looking at nameplate:
HART = -xxx5xx and -xxx6xx
Upgrade to LSCH-H1 / -H2
– No or wrong DD
(device description)
For further information see chpt. 6,
“Interfaces”
– HART interface missing
– Instrument not registered with
HART server
– Load < 230
Ω
– HART receiver (e.g. FXA 191) not
connected via load
– Incorrect device address
(addr. = 0 for single operation,
addr. > 0 for multi-drop operation)
– Line capacitance too high
– Line interference
No Profibus PA
communication
– No central PA module
Verify by looking at nameplate:
Profibus PA = -xxx3xx
Upgrade to LSCP module
– Wrong SW version (without PA)
For further information see chpt. 6,
“Interfaces”
Notes on project planning for
Profibus PA can be found in
Technical Information TI 260F,
detailed information on
instrumentation and accessories in
the operating instructions BA 198F.
– Commuwin (CW) II:
CW II version and instrument
SW version incompatible
– No or wrong DD/DLL
– Incorrect baud rate setting for
segment coupler in DPV-1 server
– Incorrect station (master)
address or duplicate address
– Incorrect station (slave) address
– Bus line not terminated
– Line problems (too long, cross
section too small;
not shielded, screen not
grounded, wires not twisted)
– Bus voltage too low (bus supply
voltage typ. 24 V DC for non-Ex,
13.5 V DC for Ex)
Voltage at instrument PA
connector must be at least 9 V.
PM253E07.CHP
Liquisys M CPM 223 / 253
Maintenance and troubleshooting
Hauser
55