5.2.2
Setup 2
Coding
Field
Selection or range
Factory setting
Display
Info
B
Function group
SETUP 2
Initial display in function group
SETUP 2.
B1
Selection of type
of temperature
compensation
(for process)
Operation mode pH:
1
ATC
MTC
Operation mode ORP:
Off
On
If B1 = ATC, the display jumps
to B3.
If B1 = MTC, enter the process
temperature in B2 which is to be
used for compensation.
B2
Enter process
temperature
25.0 °C
–20.0 ... 150.0 °C
Only if A1 = pH and B1 = MTC.
The display value can be edited.
The setting can only be
performed in °C.
B3
Selection of type
of temperature
compensation (for
calibration)
ATC
MTC
If B3 = ATC, the display jumps
to B5.
If B3 = MTC, enter the
compensated temperature in B4.
A separate temperature
sensor must also be
immersed in the buffer
solution.
B4
Enter corrected
process
temperature
25.0 °C
–20.0 ... 150.0 °C
Only if B1 = ATC and B3 = ATC.
The display value can be edited.
B5
Temperature
difference (offset)
is displayed
0.0 °C
–5.0 ... 5.0 °C
Only if B1 = ATC.
The difference between the
temperatures measured and
entered is displayed.
Factory settings are printed in
bold
face;
base version does not include functions in
italics.
B
SETUP 2
ATC
B1
°C-Proces
25.0
B2
°C
MTC temp.
ATC
B3
°C-Cal
25.0
°C
B4
RealTemp
0.0
°C
B5
Temp.Offs
Instrument configuration
Liquisys M CPM 223 / 253
26