5.3
Current outputs
The function group CURRENT OUTPUT is
used to configure the individual outputs.
Either a linear (O2 (1)) or, in conjunction with
the S version, a user-defined current output
characteristic (O2 (3)) can be entered.
Furthermore, a current output value can be
simulated to check the current outputs (O2 (2)).
Coding
Field
Selection or range
Factory setting
Display
Info
O
Function group
CURRENT
OUTPUT
Initial display in function group
CURRENT OUTPUT.
O1
Selection of
current output
Out1
Out2
A different characteristic can be
selected for each output.
O2 (1)
Selection of linear
characteristic
lin
=
linear
(1)
sim = simulation (2)
table (3)
The slope of the characteristic
may be positive or negative.
See O2 (2), O2 (3) for the
simulation and table
characteristics.
O211
Selection of
current range
4–20 mA
0–20 mA
O212
0/4 mA value;
enter
corresponding pH
(redox) or
temperature value
pH 2.00
pH –2.00 ... 16.00
–1500 mV
–1500 ... 1500 mV
0.0 %
0.0 ... 100.0 %
0.0 °C
–20.0 ... 150.0 °C
Enter the measured value
corresponding to the minimum
current value (0/4 mA) at the
transmitter output. (Spreading:
see Techn. data on p. 68.)
O213
20 mA value;
enter
corresponding pH
(redox) or
temperature value
pH 12.00
pH –2.00 ... 16.00
1500 mV
–1500 ... 1500 mV
100.0 %
0.0 ... 100.0 %
100.0 °C
–20.0 ... 150.0 °C
Enter the measured value
corresponding to the maximum
current value (20 mA) at the
transmitter output. (Spreading:
see Techn. data on p. 68.)
O2 (2)
Current output
simulation
lin
= linear (1)
sim
=
simulation
(2)
table (3)
The simulation is terminated by
selecting (1) or (3).
See O2 (1), O2 (3) for other
characteristics.
Factory settings are printed in
bold
face;
base version does not include functions in
italics.
O
OUTPUT
Out1
O1
Sel.Out
lin
O2
Sel.Type
4-20
O211
Sel.Range
2.00
pH
O212
0/4 mA
12.00
pH
O213
20 mA
sim
O2
Sel.Type
PM253E5A.CHP
Liquisys M CPM 223 / 253
Instrument configuration
Hauser
27