Coding
Field
Selection or range
Factory setting
Display
Info
R257
Enter pause time
360 min
1 ... 7200 min
The pause time is the time
between two cleaning cycles
(see chpt. 5.5.3, p. 36).
R258
Enter minimum
pause time
120 min
1 ... R257 min
The minimum pause time
prevents continuous cleaning
when the cleaning trigger is
present.
R259
Enter number of
cleaning cycles
without cleaning
agent (economy
function)
0
0 ... 9
Cleaning with cleaner can be
followed by up to 9 cleaning
cycles that use only water until
the next cleaning cycle with
cleaner is performed.
R2 (6)
Configure
neutralisation
controller
(S version only;
assignment of
contacts 1 and 2)
LC PV =
limit contactor
pH/redox (1)
LC °C =
limit contactor T (2)
PID controller (3)
Timer (4)
Clean =
Chemoclean (5)
Neutr
=
Neutra controller
(6)
Only if A1 = pH.
If the neutralisation controller is
selected for Rel1, only the neutra
controller option is offered for
Rel2.
R261
Switch function of
R2 (6) off or on
off
on
R262
Enter set point 1
(or 2)
pH 6.00
pH –2.00 ... 16.00
R263
Enter control gain
K
p
1 (or K
p
2)
1.00
0.10 ... 20.00
R264
Enter integral
action time T
n
(0.0 = no I comp.)
0.0 min
0.0 ... 999.9 min
Factory settings are printed in
bold
face;
base version does not include functions in
italics.
360
R257
PauseTime
min
120
R258
Min.Pause
min
0
R259
EconomyCl
Neutr
R2
Sel.Type
off
R261
Function
6.00
R262
pH
Setpoint1
1.00
R263
Kp Rel1
0.0
R264
min
Tn Rel1
PM253E5B.CHP
Liquisys M CPM 223 / 253
Instrument configuration
Hauser
43