Coding
Field
Selection or range
Factory setting
Display
Info
O221
Enter simulation
value
current value
0.00 ... 22.00 mA
The current value entered here is
output through the current output.
O2 (3)
Enter current
output table
(S version only)
lin
= linear (1)
sim = simulation (2)
table
(3)
Values may also be added or
changed at a later point in time.
The values entered are auto-
matically sorted in ascending
order by current value.
See O2 (1), O2 (2) for other
characteristics.
O231
Selection of table
option
read
edit
O232
Enter number of
value pairs
1
1 ... 10
.
O233
Selection of table
value pair
1
1 ... number of table
value pairs
asign
O234
Enter x value
pH 0.00
pH –2.00 ... 16.00
0 mV
–1500 ... 1500 mV
0.0 %
0.0 ... 100.0 %
x value = measured value
determined by user.
O235
Enter y value
0.00 mA
0.00 ... 20.00 mA
y value = current value
determined by user to be
associated with x value in O234.
O236
Enter whether or
not the table
status is okay
yes
no
Return to O2.
O2 (4)
Return to O1
lin
= linear (1)
sim = simulation (2)
table (3)
Factory settings are printed in
bold
face;
base version does not include functions in
italics.
10.20
O221
mA
Simulat.
O2
Sel.Type
table
O231
Sel.Table
read
O232
No.Elem.
1
O233
Sel.Elem.
1
O234
Meas.val.
0.00
pH
O235
mA value
0.00
mA
O236
Status ok
yes
O2
Sel.Type
Instrument configuration
Liquisys M CPM 223 / 253
28
Hauser