Connecting the analyzer
NOTICE
The device does not have a power switch
‣
You must install the device near (distance < 3 m (10 ft)) an easily accessible and fused plug
socket so that it can be disconnected from the power supply.
Compliance with the protective grounding specifications is mandatory when installing
the device.
5.1.1
Cable types
Analog, signal and transmission cables: e.g. LiYY 10 x 0.34 mm
2
5.1.2
Routing the cables
1. Remove the bottle tray (2) by lifting up the recessed grip slightly and then pulling it
towards the front.
2
A0028956
2
Bottle tray
2. Using a Torx screwdriver (T25), release the five screws on the carrier board (3) and fold
out the board towards the front.