Liquiline System CA80AL
Diagnostics and troubleshooting
Hauser
127
No.
Message
Factory settings
Tests or remedial action
S
D
F
732
Wear parts alarm
F
On
On
One or more wear parts has exceeded its service
life.
‣
Replace the wear parts in question and reset
the counter in
Diagnostics
/
Operating time
information
.
733
Wear parts warning
M
On
Off
One or more wear parts has almost reached the
end of its service life.
‣
Replace the wear parts in question and reset
the counter in
Diagnostics
/
Operating time
information
.
906
Cat.exchanger failure
F
On
Off
Invalid values for conductivity or flow
1.
Check for valid measured values in the
menu of the mathematical function.
2.
Check sensors.
3.
Check minimum flow.
907
Cat.exchanger warning
S
On
Off
Limit values exceeded for conductivity or flow.
Possible reasons:
• Resin depleted
• Blocked pipe
‣
Check application.
908
IEX capacity low
M
On
Off
The capacity of the exchange resin will soon be
exhausted.
‣
Schedule resin regeneration or replacement.
909
IEX capacity exhausted
F
On
Off
The capacity of the exchange resin is exhausted.
‣
Regenerate or replace the resin.
910
Limit switch
S
On
Off
Limit switch activated
930
No sample
F
On
On
Sample flow interrupted during aspiration
• Suction line blocked or leaking
• No inflow of sample
1.
Check the suction line and suction strainer.
2.
Check inflow of sample.
931
Sampling time
M
On
Off
Standard pumping time exceeded
• Filter cartridge clogged
• Sample line partially blocked
1.
Clean the filter cartridge.
2.
Clean the sample line.
3.
Replace the filter or sample line.
932
Cleaning failure
M
On
On
Cleaning failed
• Suction line blocked or leaking
• Compressed air supply defective
• No inflow of cleaner or sample
1.
Check the suction line.
2.
Check the compressed air supply and hoses.
3.
Check the cleaner and cleaner pump.
4.
Check inflow of sample.
936
Temp. range exceeded
S
On
Off
Outside temperature for sample preparation
system out of specification
1.
Check application.
2.
Check temperature sensor.
3.
Check the configuration of the temperature
sensor, sample preparation system and the
heating.