Operation
Liquiline System CA80AL
72
Hauser
Menu/Setup/Analyzer/Measurement
Function
Options
Info
Date
01.01.1970 to 07.02.2106
Factory setting
DD.MM.YYYY
Time
00:00:00 to 23:59:59
Factory setting
HH:MM:SS (24 h)
Measuring interval
Factory setting
0:10
For configuring the time interval of the
measurement
Sequence of measurements
Selection
• SP1
• SP2
• SP3
• SP4
• SP5
• SP6
• Pause
Defines the sequence of channels in automatic
mode.
Is displayed only in the case of a two-channel
device.
•
SP1
: Indicate the number of consecutive
measurements for channel
SP1
•
SP2
: Indicate the number of consecutive
measurements for channel
SP2
•
Pause
: If the
Pause
channel is selected, the
analyzer does not run the measurement.
Using the soft keys
INSERT
,
DEL
and
SAVE
you
can add rows to the table, delete and save them.
Channel change delay
15 to 600
Factory setting
30
Additional delay time which the analyzer waits
to elapse after every change of sample.
Signal delay
0 to 600
Factory setting
0
Here you can delay the process of outputting the
"Measurement active" signal when a
measurement is running by the specified time.
Measuring is paused for the duration of the
defined signal delay time.
10.3.3 Calibration
Menu/Setup/Analyzer/Calibration
Function
Options
Info
Start condition
Selection
• Immediate
• Date/time
Factory setting
Immediate
The calibration can start either immediately or
on the set date/at the set time.
If the start condition selected is
Immediate
Date
01.01.1970 to 07.02.2106
Factory setting
DD.MM.YYYY
Time
00:00:00 to 23:59:59
Factory setting
HH:MM:SS (24 h)
Calibration interval
0-01 to 90-00 (DD-HH)
Factory setting
02-00
For configuring the time interval of the
calibration/adjustment.
Next calibration
Mode
=
Automatic
Read only
Zero point
Read only