Liquiline System CA80AL
Operation
Hauser
69
Menu/Setup/Analyzer/Extended setup
Function
Options
Info
Signal for sample request
Lead time SP%C
Selection
0.00 to 30.00 (MM:SS)
Factory setting
0.00 (MM:SS)
In the automatic mode, each measurement only
starts once the lead time has elapsed. The
request sample signal is active from the start of
the lead time until the end of the specified signal
duration time.
Duration SP%C
Selection
0:00 to 60:00 (MM:SS)
Factory setting
3:00 (MM:SS)
You can set the duration for which the signal is
active.
Signal for process access
SP1
Selection
• Always accessible
•
Binary input
x:y
Factory setting
Always accessible
Always accessible
: The input signal level at the
binary inputs does not affect actions that require
sample (measurement, calibration, cleaning).
Binary input
x:y: When the signal is active at
the selected input, the analyzer starts actions
where sample is required. Otherwise the
analyzer defers or skips the action.
In the case of two-channel devices
SP2
is also
displayed.
Diagnostics settings
Limits dispensers
Control
Selection
• Off
• On
Factory setting
On
Warning limit
Remaining
operating time
Read only
Dispenser 2
,
Dispenser 3
,
Dispenser 4
,
Dispenser 5
,
Dispenser 7
Selection
1 to 90 (d)
Factory setting
28 d
Diag. code 733
Read only
Alarm limit
Remaining
operating time
Selection
1 to 60 (d)
Factory setting
7 d
Diag. code 732
Read only
Bottles
Control
Selection
• Off
• On
Factory setting
Off
Bottle filling levels
Start volume