Electrical connection
Liquicap M FMI52 HART
20
Hauser
• 12.0 to 36.0 V
DC
in the non-hazardous area
• 12.0 to 30.0 V
DC
in the Ex ia hazardous area
• 14.4 to 30.0 V
DC
in the Ex d hazardous area
5.2
Wiring and connecting
5.2.1
Connection compartment
Depending on explosion protection, the connection compartment is available in the following
variants:
Standard protection, Ex ia protection
• polyester housing F16
• stainless steel housing F15
• aluminum housing F17
• aluminum housing F13 with gas-tight process seal
• stainless steel housing F27
• aluminum housing T13, with the separate connection compartment
Ex d protection, Gas-tight process seal
• aluminum housing F13 with gas-tight process seal
• stainless steel housing F27 with gas-tight process seal
• aluminum housing T13, with the separate connection compartment
Connecting the electronic insert to the power supply:
1.
2.
3.
4.
A0040635
1. Unscrew the housing cover.
2. Remove the housing cover.