Manufacturer's Declaration
Proline Promass 500
6
Hauser
Analysis of HW/SW field data for evidence of "prior use" according to IEC
61511
Assessment by FMEDA according to IEC 61508-2 for devices without software
Assessment by (including report
no. + FMEDA data source)
TÜV Rheinland Industrie Service GmbH – Certificate No. 968/FSP 1407.00/17
Test documents
Development documents, test reports, data sheets
1)
No continuous operation as per IEC 61508: 2011 (section 3.5.16)
SIL integrity
Systematic safety integrity
SIL 2 capability
SIL 3 capability
Hardware safety integrity
Single-channel service
(HFT = 0)
SIL 2 capability
SIL 3 capability
Multi-channel service
(HFT ≥ 1)
SIL 2 capability
SIL 3 capability
FMEDA
Safety function(s)
Min., Max., Range
Device model
A1
A2
Option BA, BB
Option CA, CB
Option BA, BB
Option CA, CB, CC
λ
DU
127 FIT
127 FIT
145 FIT
146 FIT
λ
DD
1480 FIT
1439 FIT
2131 FIT
2172 FIT
λ
SU
994 FIT
957 FIT
1172 FIT
1209 FIT
λ
SD
1319 FIT
1348 FIT
2131 FIT
2102 FIT
SFF - Safe Failure Fraction
97 %
97 %
PFD
avg
for T
1
= 1 year
(single-channel architecture)
5.6 · 10
-4
6.3 · 10
-4
PFD
avg
for T
1
= 4 years
(single-channel architecture)
2.2 · 10
-3
2.5 · 10
-3
PFH
6.4 · 10
-8
6.3 · 10
-8
7.3 · 10
-8
PTC
Up to 99 %
Up to 99 %
MTBF
tot
56 years
58 years
47 years
48 years
Diagnostic test interval
30 min
30 min
Fault response time
30 s
30 s
Process safety time
50 h
50 h
Содержание HART Proline Promass 500
Страница 2: ......
Страница 8: ...Certificate Proline Promass 500 8 Endress Hauser 2 Certificate A0033748 ...
Страница 50: ......
Страница 51: ......
Страница 52: ...www addresses endress com ...