Operation
DTT31, DTT35
18
Hauser
5.1.6
Basic settings
5.1.7
Calibration
Function
group
Function
Set-
tings
Description
Basic settings
Display
66
66
PV: displays the current measured value
PVRO: displays the current measured value
rotated 180°
TMP: displays the current medium temperature
TMPR: displays the current medium temperature
rotated 180°
OFF: display off
OFFR: display off, rotated 180°
Factory setting:
current measured value (PV)
Technical
unit
C
C6
Display medium temperature unit °C or °F
Factory setting:
°C
!
Note!
Only visible if the current medium temperature
TMP is selected in the DISP mode.
Damping
,M,
Measured value damping with regard to display
value and output: 0 (no damping) or 9 to 40 s (in
increments of 1 second)
Factory setting:
0 s
DESINA
Only for 2
x PNP
switch out-
puts
Behavior as per DESINA: PIN assignment of the
M12 connector is in accordance with the guide-
lines of DESINA
(DESINA = DistributEd and Standardized INstAl-
lation technology for machine tools and manufac-
turing systems)
Factory setting:
NO
Function
group
Function
Set-
tings
Description
Calibration
6
Learn High
Flow
RUN, WAIT: setting for the maximum flow rate
occurring. 100% value (
→
Page 14)
6
Learn Low
Flow
RUN, WAIT: setting for the minimum flow rate
occurring. 0% value (
→
Page 14)