DTT31, DTT35
Troubleshooting
Hauser
29
8
Troubleshooting
8. 1 Error messages and warning messages
If an error occurs in the device, the color of the status LED changes from green to red and the
illumination of the digital display changes from white to red. A status LED flashing red and green
signals a warning. The display shows:
• E-code for errors
In the event of an error message, the measured value is uncertain.
• W-code for warnings
In the event of a warning, the measured value is reliable.
Code
Explanation
Remedy
E011
Device configuration faulty
Reset device (
E012
Error in measurement or medium temperature
outside measurable range
Check medium temperature, return
device to Hauser where
necessary
E013
Sensor heating defective
Return device to Hauser
E019
Power supply outside specification
Check operating voltage
E015
Memory error
Return device to Hauser
E020
E021
E022
Power is only supplied to the device via the
communication interface (measurement is deac-
tivated)
Check operating voltage
E042
Output current can no longer be generated
(only for 4 to 20 mA output, e.g. load at analog
output too high or open analog output)
Check load; switch off analog output