3 Installation
CleanFit P CPA 471
10
Hauser
3.2.2 Installation location/position
The CPA 471 assembly is designed for installation on tanks or pipes. Suitable
connections must be available.
!
Note!
It is imperative to maintain an installation angle of min. 15°.
Fig. 3.6:
Installation examples: The minimum installation angle is 15° to the horizontal.
!
Note!
•
The assembly requires a minimum nominal width of DN 80 for installation in a pipe.
•
For thinner pipes please use a flow vessel. This is available as an accessory (DN 25,
stainless steel SS 316L; Order No., see Accessories section).
Fig. 3.7:
Flow vessel for CPA 471; DN 25, stainless steel 316L
"
Caution!
Risk of frost damage!
If there is a risk of temperatures below 0°C, the assembly and any air or water tubes must be
heated.