1 Introduction
Applications
The Cerabar S pressure transmitter measures the pressure of gases, steam/vapour and
liquids and can be used in all areas of chemical and process engineering.
Operating principle
Metal sensor
The
process
pressure
deflects
the
separating diaphragm and a fill fluid
transmitts the pressure to a resistance
bridge. The bridge output voltage, which is
proportional to pressure, is then measured
and processed.
Level measurement
The hydrostatic pressure of a column of
liquid enables its level to be measured
continuously by a pressure transmitter if the
density
ρ
of the liquid is known.
0 - 10 bar
Z
S
KTA 3505
IEE 323/344
Version- x
Dat./Insp.:
HAUSER
CERABAR S
Order Code: PMP71K-xxxxxxxxxx
Ser.-No.: xxxxxxxxxxx
p 0…10 bar
Pmax 26.7 bar
U 45 VDC
Mat 1.4435/316L
Span 0.011/10 bar
4…20 mA
1.4435/316L 1.4435/316L
IP 65
Cal./Adj.
TAG #
BA234Y03
Figure 1.1
Examples of the Cerabar S
pressure transmitter
BA2
34Y04
Resistance
bridge
Fill fluid
Diaphragm
Metallic measuring cell
Figure 1.2
Ceramic and metallic measuring cell
h =
p
hydr
ρ
•
g
1 Introduction
Cerabar S
6
Hauser