Description of
parameters
(continuation)
Parameter
Description
Serial No. Sensor
(VAH3)
Indicates the serial number of the sensor.
Process
Connection P+
(VAH4)
Select and display the process connection material on the plus side.
Options: steel, 304 stainless, 316 stainless, Hastelloy C, Monel, tantalum, titanium,
PTFE (Teflon), 316L stainless, PVC, Inconel, ECTFE and special (special versions)
Process
Connection P–
2)
(VAH5)
Select and display the process connection material on the minus side.
For options, see parameter "Process Connection" (VAH4).
Gasket
(VAH6)
Select and display the gasket material.
Options: FPM Viton, NBR, EPDM, urethane, IIR, KALREZ, FPM Viton for oxygen
applications, CR, MVQ and special (special versions).
Process Diaphragm
(VAH7)
Select and display the diaphragm material.
Options: 304 stainless, 316 stainless, Hastelloy C, Monel, tantalum, titanium, PTFE
(Teflon), ceramic, 316L stainless, Inconel, special (special versions).
Fill Liquid
(VAH8)
Select and display the oil filling.
Options: silicon oil, vegetable oil, glycerine, inert oil, HT oil (high-temperature oil),
special (special versions).
2) These parameters are only relevant for diffenetial pressure transmitters.
Cerabar S
10 Operating Matrix
Hauser
59