Liquiline System CA80COD
Electrical connection
Hauser
27
85
85 86
86
41
43
42
87
88
97
98
31
31
32
32
SD
Display
1
2
Sensor
0/4 ... 20 mA
Sensor supply
+
+
–
+
+
–
1
2
Service
PK
GY
PK
GYGY
1
2
–
BN
WH
GN
YE
BN
WH
GN
YE
87
88
97
98
+
–
A
B
+
–
A
B
–
Alarm
HART
Power
(internal)
24VDC
24V
12,5V
3,3V
A0016537
15
E basic module wiring diagram
5.2.2
Connecting the sensors
Only use terminated genuine cables where possible.
GN/YE
YE
GN
BN
WH
GND
+
–
Com A
Com B
GY
A0024019
16
Example of Memosens data cable CYK10
Connecting the ferrules of the sensor cable to the E basic module
1. To access the electronics compartment, proceed as described in the "Routing the cables"
section.
2. Guide the sensor connecting cable from below through the cable gland on the inner rear
panel of the device and feed it upwards into the electronics compartment.
3. Establish the connection as per → 16, 27