Commissioning
Hauser
43
L
CAUTION
Analyzer in operation and during maintenance activities
Risk of injury and infection from medium or reagents
‣
Before hoses are released make sure that no action, such as the pumping of sample, is
currently running or is due to start soon.
‣
Wear protective clothing, goggles and gloves or take other suitable measures to protect
yourself.
‣
Wipe up any escaping reagent using a disposable cloth and rinse with clear water. Then dry
the cleaned areas with a cloth.
‣
Self-priming: Connect the intake hose supplied (3 m) to the peristaltic pump ("sample") (see
hose connection diagram), and guide the hose through the analyzer's hose gland to the
outside.
Ensure that only sample that is aqueous and homogenized is supplied, as otherwise
there is a risk of blockage. The customer must guarantee a constant and sufficient
volume of sample.