Liquiline System CA80COD
Commissioning
Hauser
45
L
WARNING
Connection errors
The safety of people and of the measuring point is under threat. The manufacturer does not
accept any responsibility for errors that result from failure to comply with the instructions in
this manual.
‣
Put the device into operation only if you can answer
yes
to
all
the following questions.
Instrument status and specifications
‣
Are the hoses free from damage on the outside?
Pressure reactor
‣
Have all of the reactor's connections been installed correctly?
‣
Has the reactor's safety cover been installed?
Visual inspection of the liquid-bearing lines
‣
Is the suction line connected to the peristaltic pump?
‣
Is the dispenser correctly inserted?
‣
Can the dispenser move freely?
‣
Are all the hose connections leak-tight?
‣
is the sample hose in the hose gland strain-relieved?
‣
Have the bottles with reagents, and standard been inserted and connected?
‣
Inspect the hose connections. Use the hose connection diagram as a guide.
7.3
Switching ON the measuring device
‣
Switch on the supply voltage. Wait for initialization.
7.4
Setting the operating language
Configure language
‣
Press the soft key
MENU
. Set your language in the top menu item.
The device can now be operated in your chosen language.
7.5
Configuring the measuring device
7.5.1
Basic setup analyzer
Making basic settings
1. Go to the menu
Setup
/
Basic setup analyzer
.
Make the following settings.
2.
Device tag
: Give your device any name of your choice (max. 32 characters).
3.
Set date
: Correct the set date if necessary.
4.
Set time
: Correct the set time if necessary.