Operation
SPECORD PLUS
24
}
In the main window, select menu item File | Close.
The systems displays the message "Shut down routine is running.".
}
After exiting the ASpect UV program, switch off the SPECORD PLUS via the mains
switch.
ü
The SPECORD PLUS is shut down.
If the measurements on the SPECORD PLUS have been completed and only the analysis
of the data in the ASpect UV remains, you can switch to the simulation mode. This cuts
the connection to the device. You can switch off the device after this and exit the ASpect
UV program later.
6.2
Switching the lamps on and off
The lamps are switched on with the device. The activation routine for the lamps can be
configured in the Options / Start / SPECORD PLUS window. The following options and
button are available:
Option
Description
[
UV off
]
/
[
UV on
]
Switches the UV lamp on or off.
[
Vis off
]
/
[
Vis on
]
Switches the VIS lamp on and off.
Switch on the UV
lamp
When active, the UV lamp is switched on with the device.
Switch on the Vis
lamp
When active, the VIS lamp is switched on with the device.
If the lamp corresponding to the lamp replacement is not lit during the start of a mea-
surement, a prompt will appear asking if the lamp should be switched on. The measure-
ment start must be repeated after this.
Switching the lamps on and off
manually
If necessary, the lamps can be switched on and off manually to extend the service life of
the lamps. Observe that the lamps require some time after being lit to reach a stable en-
ergy state. Before long-term measurements, the lamps should be switched on for at
least 2 hours to keep the drift caused by the lamps are low as possible. More informa-
tion on the service life of the lamps can be found in the section on lamp replacement (→
"Replacing lamps and dessicant cartridges"
}
Select menu item Settings | Options.
}
Click on
[
UV on
]
or
[
Vis on
]
in the Start / SPECORD PLUS tab.
The UV or VIS lamp is switched on.
}
To switch the respective lamp off, click
[
UV off
]
or
[
Vis off
]
.
The lamps are switched off.
6.3
Carrying out a first measurement with the SPECORD PLUS
The SPECORD PLUS is operated via the ASpect UV program. The program is modular and
provides measurement parameter settings and data analysis for various applications.
The description can be found in the "ASpect UV software for UV/VIS spectrophotometers"
operating manual.
A spectral measurement in the range of 250 to 700 mm is used as an example. The
holmium oxide glass filter from the Hellma standard filter set can be used as a sample, if
necessary. Measurement is carried out with the holder for standard cells.