Installation and commissioning
CyBio Carry
28
Fig. 22
Properties of <device> transfer window
}
Assign a unique name to the transfer position in accordance with the suggested
naming scheme.
}
Click on the arrow button in the
Tools
|
Default finger rocker
section to open the
context menu. Select
Teach dialog
.
}
Define the transfer position using the teach dialog (→ "Using the teach dialog"
}
When you close the teach dialog, the software takes over the X, Y and Z-coordinates
in the
Acuator positions
section and activates the tool. The entry for
Tools
|
Default
finger rocker
changes from
No
to
Yes
.
}
Click
[OK]
to close the window.
ü
You have defined a transfer position for transferring labware in the device config-
uration.
Defining a parking position
}
Define each parking position as follows:
}
Open the
CyBio Carry
window.
}
Define the parking position as a new relative position.
– Go to the
Relative positions
section.
– Click on the arrow next to
Add ...
to open the
Properties of <device> park
win-
dow.
}
Name the parking position in accordance with the naming scheme.
}
Select the correct parking position in the
General settings
|
Reference position
sec-
tion.
}
In the
Acuator positions
|
Z(mm)
section, enter a suitable height offset for the park-
ing position, e.g., 10 mm.