98
Rev B 2022/03
Troubleshooting
CyBio Well vario
No residual ejection
After aspiration of a vol-
ume, there was no residual
ejection of dosing stock dur-
ing the remaining program
sequence. On completion of
a program, there must be
no dosing liquid left in the
pipette tips.
Ad a
[Dispense with addi-
tional stroke]
command at
the appropriate point in a
sequence.
Faulty horizontal position
Command includes invalid
setting for horizontal posi-
tion.
Check your
software and hardware con-
figurations for mutual
agreement (refer to
Hard-
ware system information
window in
[Devices]
)
menu.
Faulty height
Command includes invalid
height specification.
Check your
software and hardware con-
figurations for mutual
agreement.
System not operational
System currently unable to
accept commands. If this
error occurs during a pro-
gram start, a program was
typically canceled while the
system waited for a key-
press signal.
Press a key at the CyBio
Well vario and try again.
Failure to find internal
zero-position of piston
motion drive.
System was unable to find
the zero-position for piston
motion.
Perform a change of tips.
Repair work is necessary if
this error persists.
No tip magazine available
System failed to detect a tip
magazine.
Insert a tip magazine if no
tip magazine available.
Remove and reinsert tip
magazine if a magazine had
been installed.
No or faulty communica-
tion with X-Y-stage
Data transfer to stage con-
troller interrupted.
Switch CyBio Well vario
power supply shortly off and
on again.
Pistons not in lower zero-
position
On triggering of a
[Move
piston to zero-position]
command, pistons were not
transferred to the range
below zero-position by
command
[Dispense with
overstroke]
.
Add this command at the
appropriate place in the
program sequence.
Command unavailable
Software transferred wrong
command to the system.
Check that available soft-
ware version is actually con-
figured for your system
(refer to
Hardware system
information
window in
[Devices]
) menu. If all set-
tings are found to be cor-
rect, check system
connections for proper con-
dition. If this fails to remove
the error, you should notify
the customer service.
Error message
Cause
Remedy
Содержание Analytik Jena CyBio Well vario
Страница 1: ...User Manual CyBio Well vario ...
Страница 3: ...Rev B 2022 03 III CyBio Well vario ...
Страница 5: ...Rev B 2022 03 V CyBio Well vario ...
Страница 6: ...VI Rev B 2022 03 CyBio Well vario ...
Страница 7: ...Rev B 2022 03 VII CyBio Well vario ...
Страница 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Страница 12: ...XII Rev B 2022 03 CyBio Well vario ...
Страница 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Страница 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Страница 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Страница 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Страница 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Страница 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Страница 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...
Страница 142: ...126 Rev B 2022 03 CyBio Well vario ...
Страница 146: ...130 Rev B 2022 03 CyBio Well vario ...
Страница 150: ...134 Rev B 2022 03 CyBio Well vario ...