Rev B 2022/03
87
CyBio Well vario
Operation
10. Pull magazine adapter out of pipetting head from the front. Use recessed handle at
magazine adapter bottom for this purpose.
.
Fig. 113: Display – head change – remove magazine adapter
Fig. 114: Removal of magazine adapter
(head replacement from the back version)
11. Insert tip magazine into pipetting head until mechanical stop position.
.
Fig. 115: Display – head change – insert magazine
CAUTION
There is danger of piston or system damage!
Insert only tip-filled magazines into the pipetting head! Insertion of a magazine is
monitored by a microswitch. If a magazine is found to be wrongly inserted, an "Insert
tip magazine" prompt will appear.
A microswitch failure may result in damage to the system (notably, pistons). Shut power
supply fully down and promptly notify the manufacturer in any such case.
Содержание Analytik Jena CyBio Well vario
Страница 1: ...User Manual CyBio Well vario ...
Страница 3: ...Rev B 2022 03 III CyBio Well vario ...
Страница 5: ...Rev B 2022 03 V CyBio Well vario ...
Страница 6: ...VI Rev B 2022 03 CyBio Well vario ...
Страница 7: ...Rev B 2022 03 VII CyBio Well vario ...
Страница 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Страница 12: ...XII Rev B 2022 03 CyBio Well vario ...
Страница 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Страница 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Страница 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Страница 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Страница 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Страница 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Страница 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...
Страница 142: ...126 Rev B 2022 03 CyBio Well vario ...
Страница 146: ...130 Rev B 2022 03 CyBio Well vario ...
Страница 150: ...134 Rev B 2022 03 CyBio Well vario ...