Rev B 2022/03
99
CyBio Well vario
Troubleshooting
Illegal command
Transfer cables are in a
non-conforming state.
Check transfer cables for
proper condition.
Software version not config-
ured for your system.
Check that available soft-
ware version is actually con-
figured for your system
(refer to
Hardware system
information
window in
[Devices]
) menu.
On failure to remove this
error, you should notify the
customer service.
Communication aborted
User break of currently run-
ning program
Microplate size not sup-
ported by software
System displays microplate
size that is not supported by
software.
Notify customer service and
request
software version that is
matched to this system ver-
sion.
Program memory location
not free
Reports attempted storage
of a program in a location
that is already occupied.
Delete program number,
then save a new program at
this location.
Note:
The command which relates
to this error is not sup-
ported in the given pro-
gram.
Piston drive failed to
reach upper end position
Piston motion has not
reached its upper end posi-
tion.
Remove this error by per-
forming an exchange of
tips. If the error persists,
repair work is required.
Notify customer service
Piston drive failed to
reach lower end position
Piston motion has not
reached its lower end posi-
tion.
Remove this error by per-
forming an exchange of
tips. If the error persists,
repair work is required.
Notify customer service
Sum of all volumes in a
program unequal zero
The sum of volumes aspi-
rated and ejected in a pro-
gram is unequal zero. On
adding all signed cycle val-
ues up with each other,
their sum must equal zero
within a given program.
Correct setting values
appropriately.
Sum of volumes too great
The sum of all volumes
aspirated and dispensed is
too great. Valid values are
within the range of 2
μ
l to
250
μ
l.
On adding all signed cycle
values up with each other,
this range must not be
exceeded.
Correct setting values
appropriately.
Sum of volumes before
and after block is unequal
zero
The sum of volumes aspi-
rated and ejected in a cur-
rent block (program,
combined commands,
repeat modules) is unequal
zero. On adding all cycle
values up with each other,
their sum must equal zero
within a given block.
Correct setting values
appropriately.
Error message
Cause
Remedy
Содержание Analytik Jena CyBio Well vario
Страница 1: ...User Manual CyBio Well vario ...
Страница 3: ...Rev B 2022 03 III CyBio Well vario ...
Страница 5: ...Rev B 2022 03 V CyBio Well vario ...
Страница 6: ...VI Rev B 2022 03 CyBio Well vario ...
Страница 7: ...Rev B 2022 03 VII CyBio Well vario ...
Страница 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Страница 12: ...XII Rev B 2022 03 CyBio Well vario ...
Страница 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Страница 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Страница 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Страница 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Страница 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...