7.11 Error messages by software
Error/Error code
Cause
Error correction
E 1
Tightening mechanism
defective
Contact your service partner or
contact the manufacturer.
E 2
Unknown head type
Remove the pipetting head.
Switch off the device. Insert the
pipetting head correctly. Turn on
the device and start an empty
software method for initializa-
tion. If the error still occurs, con-
tact your service partner or the
manufacturer.
E 3
Error accessing FRAM
Contact your service partner or
contact the manufacturer.
E 4
Tips are possibly not
tightened
Remove the pipette tips using the
software. Reattach the pipette
tips. If the error still occurs, con-
tact your service partner or the
manufacturer.
E 5
No tip magazine or liquid
handling adapter existent
Check the software method and -
whether there are tips or an LH
adapter on the specified position.
If necessary, check the correct fit
of the CyBio TipTray. Reattach
the pipette tips/LH adapters. If
the error still occurs, contact your
service partner or the manufac-
turer.
E 6
Head not properly inserted.
Insert the pipetting head cor-
rectly. Start an empty software
method for initialization. If the
error still occurs, contact your ser-
vice partner or the manufacturer.
E 7
Internal error
Contact your service partner or
contact the manufacturer.
E 8
The internal zero position
could not be found
Contact your service partner or
contact the manufacturer.
E 9
One of the passed firmware
parameters is outside the
permissible range or has a
wrong type
Check the value of the entered
parameter and correct it accord-
ingly. If the error still occurs, con-
tact your service partner or the
manufacturer.
E 10
The firmware command to be
executed is not implemented
Contact your service partner or
contact the manufacturer.
E 17
Piston drive error
Contact your service partner or
contact the manufacturer.
E 18
A firmware command expects
additional parameters
Contact your service partner or
contact the manufacturer.
E 19
Unknown tip magazine
Contact your service partner or
contact the manufacturer.
E 20
Execution was canceled by
operator
Restart the method.
E 24
Command not practicable at
moment
Check the software method to see
if the specifications match the
hardware used. If the error still
occurs, contact your service part-
ner or the manufacturer.
Содержание Analytik Jena CyBio FeliX
Страница 1: ...CyBio FeliX Operating instructions...
Страница 3: ...Rev B 2022 02 III CyBio FeliX...
Страница 4: ...IV Rev B 2022 02 CyBio FeliX...
Страница 8: ...VIII Rev B 2022 02 CyBio FeliX...
Страница 12: ...XII Rev B 2022 02 CyBio FeliX...
Страница 14: ...XIV Rev B 2022 02 CyBio FeliX...
Страница 24: ...10 Rev B 2022 02 Technical Specifications CyBio FeliX...
Страница 38: ...20 Rev B 2022 02 Safety instructions CyBio FeliX...
Страница 52: ...38 Rev B 2022 02 Technical description CyBio FeliX...
Страница 56: ...42 Rev B 2022 02 Packing transport and storage CyBio FeliX...
Страница 88: ...74 Rev B 2022 02 CyBio FeliX...
Страница 98: ...84 Rev B 2022 02 Maintenance Care CyBio FeliX...
Страница 100: ...86 Rev B 2022 02 Shutting Down CyBio FeliX...
Страница 106: ...92 Rev B 2022 02 Disposal CyBio FeliX...