Rev. B 2022/02
49
CyBio FeliX
Operation
7.3
Changing the pipetting head
The pipetting head can be inserted / replaced while the device is switched on or off.
Ensure sufficient acclimatization (same temperature at the place of installation of head
and device).
The seating of the head (X axis) should be in central position (can be moved manually).
The seating of the head (Z axis) should not be in its topmost position. We recommend
an offset of - 30 mm starting from the top vertical position for replacing the head.
CAUTION
Dangerous situation!
Risk of crushing as the louver
1
is opened and closed.
1 Does not apply to the version without enclosure/louver (e.g. 30-5015-500-24).
CAUTION
Dangerous situation!
Risk of crushing - the pipetting head has a mass of approx. 6 - 7 kg.
Always hold the pipetting head with both hands.
NOTE
Make sure that the cover magazine (transport protection; OL3316-11-200) was at-
tached before removing the pipetting head because the head must only be put down
when resting on this protective cover.
This is essential for preventing damage to the device!
Содержание Analytik Jena CyBio FeliX
Страница 1: ...CyBio FeliX Operating instructions...
Страница 3: ...Rev B 2022 02 III CyBio FeliX...
Страница 4: ...IV Rev B 2022 02 CyBio FeliX...
Страница 8: ...VIII Rev B 2022 02 CyBio FeliX...
Страница 12: ...XII Rev B 2022 02 CyBio FeliX...
Страница 14: ...XIV Rev B 2022 02 CyBio FeliX...
Страница 24: ...10 Rev B 2022 02 Technical Specifications CyBio FeliX...
Страница 38: ...20 Rev B 2022 02 Safety instructions CyBio FeliX...
Страница 52: ...38 Rev B 2022 02 Technical description CyBio FeliX...
Страница 56: ...42 Rev B 2022 02 Packing transport and storage CyBio FeliX...
Страница 88: ...74 Rev B 2022 02 CyBio FeliX...
Страница 98: ...84 Rev B 2022 02 Maintenance Care CyBio FeliX...
Страница 100: ...86 Rev B 2022 02 Shutting Down CyBio FeliX...
Страница 106: ...92 Rev B 2022 02 Disposal CyBio FeliX...