Instruction Manual
748441-D
April 2002
Rosemount Analytical Inc. A Division of Emerson Process Management
Maintenance and Service 5-25
Model CAT100
Oxygen Measurement combined with
infrared Channel:
1. Remove the analyzer from the enclo-
sure (see Section 5-2a, page 5-3)
2. Remove the analyzer cover (see Sec-
tion 5-4b, page 5-16)
3. Disconnect the connector P2 from the
OXS PCB (see Figure 5-16, page 5-
24).
4. Unscrew both fastening screws for
the fitting (see Figure 5-18, page 5-
25) and remove the fitting complete
with the sensor.
Figure 5-18. Oxygen Sensor with Infrared Channel
Front Panel
Gas connections
Fastening screws
Fitting