Instruction Manual
748441-D
April 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-21
Model CAT100
Mnemonic Description
Menu
Navigation
(
x
= multiple presses)
SP1 = 50
Temperature set point 1 = 50
MENU-MAX-MIN-ENTER
SP2 = 50
Temperature set point 2 = 50
MENU-MENU-MAX-MIN-ENTER
51.o.1 = 51.o.1 Output redirection = output 1 to 1 MENU-MENU-MENU-MAX- ENTER
InPt tYPE = t.c Input Type = Thermocouple
MENU
X
-ENTER-MAX-ENTER
t.c = J
Thermocouple type = J
ENTER-MAX-ENTER
dC.Pt = FFF.F Decimal Point = FFF.F
MENU
X
(rdG CnFG)-ENTER(dEC.Pt)-ENTER-MAX-ENTER
tEnP = C
Temperature Units = C
ENTER-MAX-ENTER
FLtr = 0004
Filter Constant = 4
ENTER-MAX-ENTER
Alr.1 = dSbL
Alarm 1 = disable
MENU
X
-ENTER-MAX-ENTER-ENTER
L.b.AL = dSbL Loop Break Alarm = disable
MENU
X
-ENTER-ENTER-MAX-ENTER-ENTER
Id = dSbL
ID Number = disable
ENTER-ENTER-MAX-ENTER
SELF = dSbL
Manual, Self Control = disable
MENU
X
(Out 1)-ENTER-MAX
o0L0 = 0000
Percent Low = 0
MIN/MAX-ENTER
o0HI = 0099
Percent High = 99
MIN/MAX-ENTER
CtrL = On.OF
Control Type = On, Off
ENTER-MAX-ENTER
Actn = rur5
Action Type = Reverse Acting
ENTER-MAX-ENTER
dEAd = 001.0
Dead Band = 1.0
ENTER-MAX/MIN-ENTER
rANP = dSbL
Ramp and Soak = disabled
MENU
X
(rANP SoAC)-ENTER-ENTER-MAX-ENTER
Final menu settings:
Make sure the controller is in
Run Mode
. Allow the heater to reach the 50 C set point temperature.
Enter the
Configuration Mode
again and change the following settings.
Mnemonic
Description
Menu Navigation (
x
= multiple presses)
CtrL = PId
Control Type = PID
MENU-ENTER
x
(CtrL tYPE)-ENTER-MAX-ENTER
Actn = rur5
Action Type = Reversing Action
ENTER-MAX-ENTER
AUto = EnbL
Auto PID = Enabled
ENTER-MAX-ENTER
AdPt = EnbL
Adaptive Control = Enabled
ENTER-MAX-ENTER
Antl = EnbL
Anti-integral = Enabled
ENTER-MAX-ENTER
Strt = dSbL
Start Auto PID = Disabled
ENTER-MAX-ENTER
CYCL = 0007
Cycle Time = 7 Seconds
ENTER
x
(CYCL timE)-MAX/MIN-ENTER
dPnG = 0004
Damping Factor = 4
ENTER-MAX-ENTER
rANP = dSbL
Ramp and Soak = Disable
MENU
x
(rANP SoAC)-ENTER-MAX-ENTER